Read out MFR Settings
Menu selection:
Menu D1
>Disp. Device : MFR
Press softkey „select“
Menu D2.2
>MFR – Settings
Press softkey „select“
Display
D3.6
„MFR - Settings“ shows up:
The displayed values can be scrolled by the
softkeys (
A
) „up“ and „down“ and show con-
secutively the following preset values:
Display
Parameter
adjustable Values
- Valve 1
:
value s
Runtime solenoid valve Y1
0...250 sec
- Valve 2
:
value s
Runtime solenoid valve Y2
0...250 sec
- Valve pause :
value s
Runtime pause of solenoid valves
-125..+125 sec
(- = overlap, + = pause)
- Valve F-up
:
value s
Solenoid valve follow-up time
0...250 sec
after compressor shut-down
- Comp S-up
:
value s
Opening period of both valves
to avoid a compressor start
against counter pressure
-125..+125 sec
- Runtime max :
value s
Maximum compressor runtime
0...10.000 sec
- Maint. max
:
value h
Maintenance interval
0...10.000 h
- Operating
:
value h
Total operating hours of the equipment
0...99.999 h
(for changes refer to service mode, page 26)
Read out MFR Info
Menu selection:
Menu D1
>Disp. Device : MFR
Press softkey „select“
Menu D2.2
>MFR – Info
Press softkey „select“
Display
D3.7
„MFR - Info“ shows up:
Version, Date and Time of the MFR software are
indicated.
MFR - Settings
>Valve 1 : 60s
Valve 2 : 60s
Valve Pause : 0s
Valve F_up : 60s
______________________
up
|
down
| select
MFR - Info
Software Vers. :V100
Date : 12-Jun-06
Time : 11:55:24
______________________
|
| back