14
· 202104LA090207003 · 09/02/21
14. Electromagnetic
Compatibility Information
Table 1 – Emission
PHENOMENON COMPLIANCE
ELECTROMAGNET-
IC ENVIRONMENT
RF emissions
CISPR 11 Group
1 Class B
Home Healthcare
environment
Harmon
-
ic distortion
IEC 61000-3-
2 Compliance
Home Healthcare
environment
Voltage fluc
-
tuations and
flicker
IEC 61000-3-
3 Compliance
Home Healthcare
environment
Table 2 – Enclosure Port
PHENOMENON
BASIC EMC
STANDARD
IMMUNITY
TEST LEVELS
Home Healthcare
environment
Electrostat-
ic Discharge
IEC 61000-
4-2
± 8kV contact
± 2kV, ±4kV, ±8kV,
±15 kV air
Radiated RF
EM filed
IEC 61000-
4-3
10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fileds
from RF wireless
communicati-
ons equipment
IEC 61000-
4-3
Refer to table 3
Rated power
frequency mag-
netic fields
IEC 61000-
4-8
30A/m
50Hz or 60Hz
Table 3 – Proximity
fileds from RF wireless
communications equipment
TEST FREQUEN-
CY (MHZ)
BAND
(MHZ)
IMMUNITY
TEST LEVELS
Professional
Healthcare faci
-
lity environment
385
380-390 Pulse modula-
tion 18Hz, 27V/m
450
430-470
FM, ±5kHz de
-
viation, 1kHz
sine, 28V/m
710
704-787 Pulse modula-
tion 217Hz, 9V/m
745
780
810
800-960 Pulse modulati-
on 18Hz, 28V/m
870
930
1720
1700-1990 Pulse modulati-
on 217Hz, 28V/m
1845
1970
2450
2400-2570 Pulse modulati-
on 217Hz, 28V/m
5240
5100-5800 Pulse modulati-
on 217Hz, 9V/m
5500
5785
15 Advice regarding
the disposal of waste
All the packaging is composed of materials that
pose no hazard for the environment and which can
be disposed of at your local sorting centre to be
used as secondary raw materials. The cardboard
may be disposed of in a paper recycling bin. The
packaging film must be taken to your local sort
-
ing and recycling centre.