LaMotte Tracer Total Chlorine PockeTester 1740 Manual Download Page 8

8

Recalling Stored Readings

NOTE: The HOLD symbol should not be displayed. If it is, exit the HOLD 

function by pressing the MODE button.

1. Press the CAL/RECALL button and then press the MODE/HOLD button

immediately after 

CAL

 is displayed; the location number (1 through 15)

will f ash.

2. The last stored reading taken will be displayed f rst. To advance through

the stored readings, press the MODE/HOLD button. The location number

is displayed f rst, followed by the reading stored in that location.

3. To exit the storage mode, press the CAL/RECALL button and the meter

will return to normal operation.

4. If the batteries are removed, all stored readings will be lost.

For Best Accuracy

1. Remove TRACER from sample when power is off. Even though power

is off, leaving TRACER in solution can shorten unit life.

2. Wash sample cups and caps and rinse completely before use.
3. If measuring solutions with large differences in chlorine concentration,

for example, 0.1 ppm and 5.0 ppm, use a separate sample cup for each

concentration.

4. Discard the sample cup when it becomes stained.
5. The process requires an acidic solution. The reagent tablet provides

suff cient buffering capacity to deal with alkalinity, calculated as calcium

carbonate, up to and in excess of 500 ppm. This will cover all common

water samples.

6. When the electrode is new or after dry storage, chlorine measurements

should be repeated at least twice until a repeatable reading is obtained.

7. Gently wipe the electrode surface with a paper towel between uses.
8. When measuring a lower chlorine concentration after a higher

concentration, the meter will temporarily give slightly elevated readings.

9. All oxidizing species that interfere with other chlorine methods, such as

manganese, iodine and bromine will interfere. Silver and mercuric ions

over 20 ppm will also interfere.

10. Do not touch the tablets. It can contaminate them.
11. Do not touch the sensor surface.

Summary of Contents for Tracer Total Chlorine PockeTester 1740

Page 1: ...Water Testing Leader Since 1919 Tracer Total Chlorine PockeTesterTM Code 1740 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Page 2: ...WARNING This set contains chemicals that may be harmful if misued Read cautions on individual containers carefully Not to be used by children except under adult supervision...

Page 3: ...ging the Display Temperature Units 6 Low Battery Indicator 6 Testing Conditioning the Electrode 7 Measurement 7 Storing Readings 7 Recalling Stored Readings 8 For Best Accuracy 8 Offset Adjustment Pro...

Page 4: ...buffer be added to the sample before testing SPECIFICATIONS Display Multifunction display with bar graph Operating conditions 32 to 122 F 0 to 50 C and 80 RH Chlorine range 0 01 to 10 00 ppm Total Ch...

Page 5: ...compartment cover 2 LCD Display 3 MODE HOLD button 4 CAL RECALL button 5 ON OFF button 6 Electrode Retaining Collar ring 7 Electrode Sensor Note The Electrode storage cap is not shown LCD Display 1 B...

Page 6: ...R is turned on it will automatically self calibrate and the SELF and CAL icons will be displayed These icons will disappear after the self calibration is complete The main display and bar graph will t...

Page 7: ...and the electrode must remain still and stable during the entire measurement 7 The display will f ash while readings are changing rapidly The result should be read when the display stabilizes The read...

Page 8: ...use 3 If measuring solutions with large differences in chlorine concentration for example 0 1 ppm and 5 0 ppm use a separate sample cup for each concentration 4 Discard the sample cup when it becomes...

Page 9: ...oximately 5 seconds until C_O appears at the bottom of the display 5 IMMEDIATELY press the MODE HOLD button repeatedly to increase the value or press the CAL RECALL button repeatedly to decrease the v...

Page 10: ...1 00 mL of the 10 000 ppm standard into another 100 mL volumetric f ask Dilute to the 100 mL line with deionized water Cap and mix This standard is 100 ppm 3 Pipet exactly 1 00 mL of the 100 ppm solu...

Page 11: ...splayed keep the electrode in standard and press the CAL key again until CAL is displayed then press the CAL and MODE keys again simultaneously Immediately release both keys 9 After approximately 5 to...

Page 12: ...tabs 3 Replace the four CR2032 batteries Observe polarity 4 Replace the battery compartment cap Cleaning the Electrode The TRACER Chlorine Electrode arrives in a factory cleaned condition and is read...

Page 13: ...meter to malfunction 5 Firmly tighten the electrode collar to create a seal with the rubber gasket between the electrode and the meter Do not over tighten 6 Cover the connector end of the removed ele...

Page 14: ...ment to be free of defects in parts and workmanship for 1 year from the date of shipment and the probe to be free of defects in parts and workmanship for 6 months from the date of shipment If it shoul...

Reviews: