The display may show the following messages:
E r r 1
Low battery
E r r 2
An attempt was made to calibrate the meter outside of the
allowable range
E r r 3
An attempt was made to calibrate the meter with a zero sample
E r r 4
Processing error
E r r 5
No blank reading – the meter has never been blanked for this
test factor.
E r r 6
The meter must be re-blanked and the sample rescanned
E r r 7
Call LaMotte Tech Service – meter may have to be returned for
repairs
l o w b a t t e r y
Low battery
>
Over range indicator
pq
More choices are available and can be viewed by scrolling up
and/or down through the display
Header
Identifies the current menu and information on units and reagent
systems if applicable. In the data logging mode the number of
the data point is displayed and the total number of data points in
the memory will be shown.
HEADER
M e n u T i t l e
Message or Result
Window
M e s s a g e
o r R e s u l t
Menu Selection Window
*
F i r s t C h o i c e
Footer
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
S e c o n d C h o i c e
a n o t h e r
a n d s o o n
16
DISPLAY & MENUS