n
ENTERING A NEW CALIBRATION
To enter a new calibration two reacted standards solutions of known
concentration are required: a “low standard” and a “high standard”. These
should be ready to use.
Use the
t
button to scroll to
New Calibration
and press
Q
/ENTER
button
to select.
EDIT USER TEST01
* Select Vial/WL
New Calibration
Decimal Places
Select Units
Input the concentration of the
LOW STANDARD
by using the
t
and
s
buttons to
scroll the first digit of the concentration into the first position on the display. Press
Q
/ENTER
button to select that digit (
1
for this example).
LOW STANDARD
0
______
* Continue
The number “
0
” is always the starting point for the next digit. Continue selecting
digits or a decimal point to enter the concentration (up to seven characters).
LOW STANDARD
1
0
_____
* Continue
“
1.5
” has been entered in this example. Press
Q
/ENTER
button four times to
input “
0
” as the last four digits. Pressing
Q
/
ENTER
after selecting the last digit
saves the concentration.
LOW STANDARD
1.5
0
___
* Continue
Input the concentration of the
HIGH STANDARD
by using the same method as for
the low standard.
HIGH STANDARD
0
______
* Continue
SMART2 COLORIMETER 1.07
43