10. SPECIFICATIONS
% Saturation Mode
Range
Resolution
Relative accuracy
0.00 – 200.0 %
0.1 %
± 1.5% of Full Scale
mg/L (ppm) Concentration Mode
Range
Resolution
Relative accuracy
0.00 – 20.00 mg/L or ppm
0.01 mg/L; 0.01 ppm
± 1.5% of Full Scale
Temperature
Range
Resolution
Relative accuracy
-5.0 – 105.0 °C (meter only)*
0.1 °C
± 0.5 °C
*Probe measures 0.0 – 50.0 °C
Salinity Correction
Range
Resolution
Method
0.0 – 50.0 ppt
0.1 ppt
Automatic correction after manual input
Barometric Pressure Correction (mm Hg)
Range
Resolution
Method
500 to 1499 mm Hg or 66.6 to 199.9 kPA
1 mm Hg or 0.1 kPA
Automatic correction after manual input
Automatic Temperature
Compensation
0.0 to 50.0°C
Manual Temperature
Compensation
0.0 to 50.0°C
Probe (DO/ Temp)
Galvanic / Thermistor
Probe Diameter
Body 12 mm, Cap 16 mm
Response Time
60 seconds to achieve 95% of the reading
% Saturation Calibration Points 100% in saturated air or air-saturated water.
0% in zero oxygen solution
% Saturation Calibration Limits Factory calibrated absolute value of 10.0%
and below for 0% point & 50% to 200% for
100% point.
Concentration Calibration
Window
±40% from the factory default measurement
value. Minimum reading allowed is 2.00
mg/L (ppm).
Temperature Calibration
Window
±5°C from factory default measurement