Turning the Meter On
When the meter is turned on, it will go through a series of displays that show
the set-up parameters.
For CON 6
For TDS 6
Press
ON/OFF
button to turn on the meter.
1.
First screen shows [
Con 6
] (or [
tdS 6
]) to identify the meter.
2.
Second screen shows [
C 1.0]
which is the conductivity cell constant, k.
Cell constants of 0.1, 1.0 or 10.0 can be selected. Refer to Section on
Advance Set-up. Default value is k=1.0.
3.
Third screen shows [
t 25.0 °C
] which is the Normalization Temperature.
Normalization Temperatures of 25 °C or 20 °C may be selected. Refer to
Section on Advance Set-up. Default value is 25 °C.