Summary of Contents for 4140-03

Page 1: ...e to be used as Chlorine Secondary Standards to verify the performance of the meter and are not intended to be used as calibration standards The meter is calibrated at the time of manufacture The stan...

Page 2: ...loth Press the READ button The concentration in ppm will be displayed within 2 seconds Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Push the READ button to turn the m...

Page 3: ...e a zero reading Wipe the DPD Chlorine Color Standard A 4147 E with a clean lint free cloth Press the READ button The concentration in ppm will be displayed within 2 seconds The reading should be with...

Page 4: ...he concentration will be displayed 015 Chlorine Scan Sample Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press ON to turn the meter on Press ENTER to select START Pre...

Page 5: ...o the chamber being sure to align the index line with the arrow on the meter Close the lid Repeat steps 4 6 with DPD Chlorine Color Standard B 4138 E and DPD Chlorine Color Standard C 4139 E If the re...

Page 6: ...ill be displayed Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press to turn the meter on Press ENTER to select Testing Menu Press ENTER to select All Tests Menu Press...

Page 7: ...he readings are not within the tolerance limits standardize the meter calibration by following the Calibration Procedure in the meter manual If the readings are still not within the tolerance limits t...

Page 8: ...ration will be displayed Wipe the DPD Chlorine Color Standard Blank 4136 E with a clean lint free cloth Press to turn the meter on Press ENTER to select Testing Menu Press ENTER to select All Tests Me...

Page 9: ...r Close the lid If the readings are not within the tolerance limits standardize the meter calibration by following the Calibration Procedure in the meter manual If the readings are still not within th...

Page 10: ...the dye standards being slightly different from the spectral qualities of DPD Readings of actual DPD chlorine reactions will not differ significantly between meters Contact LaMotte Company Technical...