3. MOUNTING, WIRING AND PLUMBING
3.1 Mounting
The 2020 ClearTrace Turbidity System is mounted on a 12 inch x 12 inch
panel with four ¼ inch mounting holes. Use ¼ inch bolts with washers to
mount the panel to a wall. Make sure the sample chamber is level and
plumb to insure proper water levels are maintained in the sample chamber.
It is important to note that the system must be fixed securely to the wall to
insure the chamber does not move when handled or touched.
Mounting tools and equipment:
•
½ inch Drill with drill bit for appropriate wall anchor.
•
Four ¼ diameter bolts – 1.5 inches long.
•
Four ¼ inch ID flat washers
•
Wrench to turn bolts
•
Level to insure the system is level when mounted
General Mounting Instructions:
1.
Place the 12x12 inch panel on the wall where it is to be mounted and
mark the placement of the upper left hole.
2.
Drill a mounting hole for the upper left corner with a masonry drill
bit.
3.
Mount the Turbidity System with the upper left bolt.
4.
Using a level, rotate the system until level, then mark the other three
mounting holes.
9
Figure 3.1: Wall mounting Clearances
Summary of Contents for 2020 ClearTrace Series
Page 1: ...2020 ClearTrace In line Turbidity System Code 5 0113 24V Code 5 0113 110V...
Page 2: ......
Page 35: ......