![Lake Shore Model 421 User Manual Download Page 49](http://html.mh-extra.com/html/lake-shore/model-421/model-421_user-manual_670638049.webp)
Lake Shore Model 421 Gaussmeter User’s Manual
Remote Operation
4-15
Device Specific Commands (Continued)
BRIGT?
Front Panel Display Brightness Query
Input:
BRIGT?[term]
Returned:
<brightness>[term]
Format:
n
(Refer to command for description)
FAST
Fast Data Mode Command
Input:
FAST <state>[term]
Format:
n
<state>
0 = Off, 1 = On
Remarks:
Fast Data Mode reaches data rates up to 18 readings per second via the Serial Interface with
a corresponding increase in corrected analog output. The front panel display does not
function in this mode. Refer to Paragraph 3.12.
FAST?
Fast Data Mode Query
Input:
FAST?[term]
Returned:
<state>[term]
Format:
n
(Refer to command for description)
FIELD?
Magnetic Field Reading Query
Input:
FIELD?[term]
Returned:
<field value>[term]
Format:
±nnn.nn
<field value>
Returns sign, 4 or 5 digits, and places decimal point appropriate to range.
Remarks:
Use FIELDM? to determine units multiplier and UNITS? to determine gauss or tesla units.
FIELDM?
Field Reading Multiplier Query
Input:
FIELDM?[term]
Returned:
<multiplier>[term]
Format:
a
<multiplier>
µ = micro = 10
-6
, m = milli = 10
-3
, (blank) = unity, k = kilo = 10
3
Remarks:
Used with FIELD? query.
FILT
Display
Filter
Command
Input:
FILT <state>[term]
Format:
n
<state>
0 = Off, 1 = On
Remarks:
Quiets the display reading. Refer to Paragraph 3.4.
FILT?
Display
Filter
Query
Input:
FILT?[term]
Returned:
<state>[term]
Format:
n
(Refer to command for description)