![Lake Shore 218 User Manual Download Page 88](http://html.mh-extra.com/html/lake-shore/218/218_user-manual_670641088.webp)
Lake Shore Model 218 Temperature Monitor User’s Manual
6-28
Remote Operation
LRDG?
Query Linear Equation Data for a Single Input or All Inputs.
Input: LRDG?
<input>
Returned: <Linear value>
. Format:
+/-nn.nnn[term]
Or if all inputs are queried:
<Input 1 Linear Value>,<Input 2 Linear Value>,<Input 3 Linear Value>,<Input 4 Linear
Value>,<Input 5 Linear Value>,<Input 6 Linear Value>,<Input 7 Linear Value>,<Input 8
Linear Value>
. Format:
+/-nn.nnn,+/-nn.nnn,+/-nn.nnn,+/-nn.nnn,+/-nn.nnn,+/-nn.nnn+/-
nn.nnn,+/-nn.nnn
Remarks:
Returns the linear equation data for an input.
<input>
specifies which input to query.
0
= all
inputs,
1-8
= individual input.
NOTE:
Use
0
(all inputs) when reading two or more inputs at
the maximum update rate of 16 rdg/s.
MNMX
Configure Minimum and Maximum Input Function Parameters
Input:
MNMX <input>, <source>
Returned:
Nothing.
Remarks:
Configures the minimum and maximum input functions.
<input>
Specifies input to configure (
1-8
).
<source>
Specifies input data to process through max/min.
1
= Kelvin,
2
= Celsius,
3
= sensor units,
4
= linear data.
Example: MNMX 1, 3[term] -
Input
1
min/max function processes data from input sensor units reading.
MNMX?
Query Minimum and Maximum Input Function Parameters
Input: MNMX?
<input>
Returned: <source>
. Format:
n[term]
Remarks:
Returns an input min/max configuration.
<input>
Specifies input to query (
1-8
).
<source>
Specifies input data to process through max/min.
1
= Kelvin,
2
= Celsius,
3
= sensor units,
4
= linear data.
MNMXRDG?
Query Min/Max Data for an Input
Input: MNMXRDG?
<input>
Returned: <min value>,<max value>
. Format:
+/-nn.nnn,+/-nn.nnn[term]
Remarks:
Returns the minimum and maximum input data.
<input>
specifies which input to query.
MNMXRST
Resets Min/Max Function for All Inputs
Input: MNMXRST
Returned:
Nothing.
Remarks:
Resets the minimum and maximum data for all inputs.
MODE
Configure Remote Interface Mode
Input: MODE
<mode>
Returned:
Nothing.
Remarks:
Configures the remote interface mode.
<mode>
specifies which mode to operate.
0
= local,
1
= remote,
2
= remote with local lockout.
Example: MODE 2[term] -
Places the Model 218 into remote mode with local lockout.
MODE?
Query Remote Interface Mode
Input: MODE?
Returned: <mode>
. Format:
n[term]
Remarks:
Returns the remote interface mode.
0
= local,
1
= remote,
2
= remote with local lockout.