CHAPTER 2:
System Overview
Model CRX-6.5K Probe Station
2.4.5 Back Side Voltage
Biasing
Experiments such as device characterization often require voltage biasing. The
biasing voltage can be introduced through a probe if the bias contact is available on
the top surface, but doing so prevents the use of the probe for other purposes. Voltage
biasing can be done through the sample holder if the bias contact is available on the
back side of the sample or if biasing is done directly through the substrate. An
optional coaxial sample holder and FT-BNC feedthrough and cable configuration
allows convenient back side biasing as illustrated in FIGURE 2-24.
2.4.6 Small Signal/
Low Noise DC/RF
Measurements
As measured signal magnitude decreases, environmental noise becomes more of an
issue. Proper setup of the experiment is crucial to extracting small signals from the
background noise. The Model CRX-6.5K offers several standard features and optional
configurations that can help.
2.4.6.1 Noise Isolation for Low Noise Measurements
The noise isolation features described in section 2.4.1.4 may be insufficient when
making low noise measurements. Please consider the following when setting up a
low noise experiment.
Never remove earth safety ground protection from electronic equipment.
1. Make every effort to isolate other noisy components (pumps, compressors,
switching power supplies) when they are added to the system.
2. Electrical and electronic devices are connected through the power line (mains)
even when isolated in the probe station. Care must be taken to prevent noise
from coupling through the power connection or earth safety ground.
3. AC noise can enter the measurement through electrical or magnetic coupling
even when the leads are isolated. Shielding (section 2.4.1.3) and sample isolation
(section 2.4.6.2) should also be considered.
FIGURE 2-24
Back side voltage bias with coaxial sample holder
Triaxial signal connector
Cryogenic
coaxial cable
ZN50
Probe
Sample stage
Cryogenic
coaxial cable
Probe
arm
base
Probe arm shield
To temperature
controller chassis
19-pin
connector
Thermal
anchors
Shield
Vacuum chamber base
Radiation shield stage
Coaxial
sample holder
Coaxial (BNC)
feedthrough
INSULATOR
CONDUCTOR
CONDUCTOR
Bias voltage