  g) Calculate|

a1-a2

- (b1-b2) |= e

If e is above 1mm, the instrument accuracy is out of tolerance, please contact the 

seller for servicing.   

 

5.3 Self-calibration & adjustment   

As shown in the following picture, the instrument have two calibration apertures, A means the left and right adjustment 

transversely (the error tested by item1), B means the front and back adjustment longitudinal (the error tested by item2). 

 

 

 

 

 

 

 

(1) Adjust with 3mm hexagon spanner. 

(2) (2)The adjustment of two directions will influence each other sometimes. When making the fine adjusting in left and 

right direction transversely, the front and back direction longitudinal will change possibly, also, when adjusting in front 

and back longitudinal, the left and right direction transversely will change possibly. 

(3) The adjustment of self-calibration screw could not exceed 4 circles. (Clockwise or anti-clockwise direction). 

(4) If the instrument could not adjust through self-calibration aperture, please contact the distributor for servicing. 

 

6. Demonstration

 

 

 

 

 

 

 

自校孔

自校孔

Summary of Contents for LS686

Page 1: ...series product output 3 laser lines Form two cross laser lines Compatible both indoor and outdoor application switchable between the continuous laser and pulse laser All green laser series do not have pulse function Self leveling laser flash and audible warning when beyond the self leveling range The alarming function can be shield and with tilt function The instrument is 360 degree rotatable and ...

Page 2: ...e rechargeable battery Plug the charger into the charging hole to charge the rechargeable battery box When charging the indication light is red and when finished charging the indication light turns from red to green You can also use power bank portable power to charge the rechargeable battery box Touching keypad External power supply socket Find adjusting hand wheel Round vial Laser output window ...

Page 3: ...the battery box 4 5 Adjust the adjustable supporting leg When using the instrument you could adjust 3pcs supporting foot to make the round vial in center if the laser line flash and with the buzzer sound which shows the instrument is out of self leveling range please adjust the round vial in center 4 6 Instrument rotate horizontally and fine adjusting Plug the adapter DC 6V and use Bubble in the c...

Page 4: ...ion light Flash continuously tilt mode on OFF tilt mode off Receiving mode indication light ON Receiving mode on OFF Receiving mode off Receiving mode indication Power indication light Tilt mode indication light Horizontal line controlling button Vertical line controlling button Receiving mode button Vertical line controlling button Vertical line and dot controlling button Horizontal line controll...

Page 5: ...e The receiving mode button can only be available when there is laser output if there is no laser output this button is invalid 4 11 Laser output controlling button Press the instrument output is as the following Models that end with D can output front dot Press the instrument output is as the following Models that end with D can output front dot 5 lines 3 lines Press the instrument output is as t...

Page 6: ...reference The instrument cannot enter into tilt mode in unlocking status tilt mode button is ineffective If the instrument using in tilt mode unlock it the instrument will exit from tilt mode the tilt mode LED is extinguished and the instrument enters into self leveling status If tilt mode is not needed please switch the tilt mode switch to OFF 5 Self check and calibration 5 1 horizontal line accu...

Page 7: ...he elevating or tripod and place it in the center of both staff level the bubble c Turn on all the laser lines to make the laser center dot on staff A record a1 dot value d Rotate the instrument by 180 degree make the cross dot on the staff B record b1 dot value e Move the tripod make the instrument 0 6m away from staff A and make the cross center on staff A record a2 value f Rotate the instrument...

Page 8: ...ust with 3mm hexagon spanner 2 2 The adjustment of two directions will influence each other sometimes When making the fine adjusting in left and right direction transversely the front and back direction longitudinal will change possibly also when adjusting in front and back longitudinal the left and right direction transversely will change possibly 3 The adjustment of self calibration screw could ...

Page 9: ... and any unprofessional disassembly will result in the damage of instrument Keep the instrument clean especially the laser output window glass and remove dust by the gentle operation of soft clean cloth Take the batteries out when the instrument is not is use for a long time and keep the instrument in the carrying case when it is not in use Item Parameter Laser wavelength 635nm Laser class ClassⅡ ...

Reviews: