Guidance and manufacture’s declaration
electromagnetic emissions
The DEVICE is intended for use in the electromagnetic environment specified below.
The customer or the user of the DEVICE should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The DEVICE is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
RF emissions
CISPR 11
N/A
The DEVICE is suitable for use in all establishments
domestic included and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
N/A
ELECTROMAGNETIC COMPATIBILITY
Table 1 - For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration
electromagnetic immunity
The DEVICE is intended for use in the electromagnetic environment specified below. The customer or the user of the DEVICE
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be at
least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m; 50 Hz or 60 Hz
3 A/m; 50 Hz or 60 Hz
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Table 2 - For all ME EQUIPMENT and ME SYSTEMS