Left or Right Side Ruler Mounting 

 

 

Optional Ruler Attachment 

 
The Main unit may be mounted both 
directions on the accessory attachment 
pattern for your convenience. 
 

Right and Left Side Ruler 

 

 

The Ruler may be attached either on the Right Side or Left Side by flipping the ruler 180° as shown in the 
illustration.  This may be useful for taking readings on the opposite side of the body. 

 

How to use ruler attachment for Extremity Measurements 

 

The ruler attachment may be used to assist in measurements of the upper and lower extremities.  It can be used as a 
place to hold the inclinometer and for alignment.  The Inclinometer inherently reads angles with respect to gravity, 
so the orientation of inclinometer needs to be positioned to read angles respect to the gravity.   The Inclinometer 
may be zeroed by pressing the ZERO button while the subject is in the neutral position  

 

X-Ray Film reading 

 

Place the X-Ray film on the lighted box mounted on the wall.  Place the ruler in line of reference position and then 
press 

ZERO

 to initialize reference angle. 

 
Then move the ruler to the second angle to be read and press 

HOLD

.  The difference angle will be stored and can be 

viewed by pressing 

VIEW

 
 