20-18071
Rev A
Replacing Batteries 2/2
Ensure the Patient Simulator is switched off.
1. Unzip the skin at shoulders and torso and place to one
side.
Note
Ensure the LiveShock cable has been disconnected.
2. Move the stomach foam carefully to one side to avoid
pulling on connecting tubes and cables.
3. Remove battery cover.
4. Disconnect the battery cable from the torso cable as
shown.