Function Test
Function Test
22
23
Inspect and test valve function to ensure proper operation of
the Laerdal Silicone Resuscitator prior to patient use. To ensure
proper operation, test valve functions after cleaning, disinfection
and reassembly.
Laerdal Silicone Resuscitators should be function tested at least
once every year.
Caution
If a Laerdal Silicone Resuscitator fails any of these function tests it is
to be correctly cleaned, re-assembled and tested or removed from
service and not used.
Testing the Intake Valve
Squeeze
Release bag
Inspect Bag - must
expand
Seal neck
Try to squeeze
Inspect Bag - no air
should escape