background image

Laerdal Silicone  

Resuscitators

User Guide

EN

www.laerdal.com

Summary of Contents for LSR Adult

Page 1: ...Laerdal Silicone Resuscitators User Guide EN www laerdal com...

Page 2: ...ssembly and Disassembly 10 Paediatric Model Overview 12 Paediatric Model Assembly and Disassembly 14 Preterm Model Overview 16 Preterm Model Assembly and Disassembly 18 IntakeValve Assembly and Disass...

Page 3: ...e product only as described in this User Guide Warning and Cautions A Warning states a condition hazard or unsafe practice that can result in serious personal injury or death A Caution states a condit...

Page 4: ...icone Mask Child 3 4 with Multi Function Mask Cover Silicone Mask Adult 4 5 with Multi Function Mask Cover Silicone Mask Size 00 0 1 2 PretermVentilation Bag 240 ml Reusable Oxygen Reservoir Bag 600 m...

Page 5: ...w Adult Model Overview IntakeValve PatientValve Reservoir FlapValves x2 Disk Membrane LipValve Silicone Mask with Multi Function Mask Cover Oxygen Port Reusable Oxygen Reservoir Bag AdultVentilation B...

Page 6: ...embrane and lip valve may affect performance Ensure use of one lip valve Misassembly with two lip valves may cause inadvertent Positive End Expiratory Pressure PEEP or prevent proper patient exhalatio...

Page 7: ...l Overview Pressure ReliefValve IntakeValve PatientValve Reservoir FlapValves x2 Disk Membrane LipValve Silicone Mask Child 3 4 with Multi Function Mask Cover Reusable Oxygen Reservoir Bag PaediatricV...

Page 8: ...ake membrane disk membrane and lip valve may affect performance Ensure use of one lip valve Misassembly with two lip valves may cause inadvertent Positive End Expiratory Pressure PEEP or prevent prope...

Page 9: ...erview Preterm Model Overview Silicone Mask IntakeValve PatientValve Reservoir FlapValves x2 Disk Membrane LipValve PretermVentilation Bag Reusable Oxygen Reservoir Bag Pressure ReliefValve Oxygen Por...

Page 10: ...intake membrane disk membrane and lip valve may affect performance Ensure use of one lip valve Misassembly with two lip valves may cause inadvertent Positive End Expiratory Pressure PEEP or prevent pr...

Page 11: ...takeValve Assembly Disassembly IntakeValve Cap Update 20 21 New version of cap and valve are indentified by textured circle Post 2015 cap Post 2015 cap Post 2015 valve W ider Pre 2015 cap Pre 2015 cap...

Page 12: ...ing disinfection and reassembly Laerdal Silicone Resuscitators should be function tested at least once every year Caution If a Laerdal Silicone Resuscitator fails any of these function tests it is to...

Page 13: ...atientValve Testing the PatientValve Caution Ensure that a single Lip Valve has been installed in the Patient Valve Inspect LipValve movement Attach reservoir bag to PatientValve Squeeze Fill reservoi...

Page 14: ...mbient Air Inspect air flow and Inlet Reservoir FlapValve movement Seal oxygen port and squeeze the bag multiple times to empty the reservoir Release Bag Attach reservoir bag to PatientValve Attach re...

Page 15: ...or The use of cleaning and disinfection procedures not described in this section may have adverse effects on the LSR material and or performance The LSR must be high level disinfected before first tim...

Page 16: ...e any further Washing and Rinsing The LSR must be cleaned before high level disinfection or sterilisation The LSR can be manually cleaned or cleaned in an automatic washer disinfector Manual Cleaning...

Page 17: ...ation cycle 70 75 C 158 167 F 30 min Dry the components thoroughly Reassembly Reassemble resuscitator as shown in Assembly Disassembly section The product is in compliance with the essential requireme...

Page 18: ...ithout reservoir 20 x 40 20 x 60 150 x 20 150 x 30 300 x 12 300 x 24 3 100 97 100 97 98 56 78 57 85 48 56 46 8 100 100 100 100 100 70 100 70 100 58 100 57 15 100 100 100 100 100 82 100 83 100 71 100 7...

Page 19: ...ng parts must be replaced and disposed of Accessories Catalogue Description 871000 Silicone extension tube 28 cm 850900 LSR Manometer connector 850500 Expiration diverter OD 30 mm 531907 LSR IntakeVal...

Page 20: ...set of 4 871950 FlapValves 2 pk 870708 Disposable O reservoir bag 2600ml pkg 20 851250 PatientValve w 35cm H O Pressure Relief Valve 521100 Wall Mount Paediatric Preterm Display Case Masks Catalogue D...

Page 21: ...cal AS All rights reserved Manufactured in China for Laerdal Medical AS P O Box 377 Tanke Svilandsgate 30 4002 Stavanger Norway T 47 51 51 17 00 Distributed in US by Laerdal Medical Corporation 167 My...

Reviews: