laerdal 846060 User Manual Download Page 2

2.b

2.c

6

1

2.a

3

4

5

5.b

5.a

1

6

5.c

4

3

2

 

Steam 136 °C
10-20 minutes

100 °C

10-20 minutes

5.b

5.a

3

Disassembly

Overview

Items Included

Assembly Highlights

Upright with PEEP 

Newborn Bag-Mask

Reprocessing Instructions

Pull Out

Pull Out

Pull Off

Lift Out

Pull firmly on tab to release

To disassemble follow steps 1-6. 

Unscrew

Unscrew

Unscrew

       Caution

Do not disassemble parts beyond the steps shown. 

1. Lip Valve Membrane Function

2. Pressure Release Valve & Inlet Valve Function

3. Product Sealing

Squeeze the bag. Check that the yellow valve opens 
and closes with every squeeze.

Seal the mask with a hand. Squeeze the bag forcefully. 
Check that air is released from the pressure release valve. 
Keep the mask sealed against hand. Release the squeezed 
bag. Check that the bag re-expands without resistance.

Keep the mask sealed against the hand. 
Press the pressure release valve down.
Squeeze the bag and check that there is no 
leakage.

Hold Upright firmly at the neck. Press other hand 
upwards against the mask’s sealing cuff so that air 
inside the mask is pushed into Upright and out the 
expiration port with the PEEP Valve membrane. 
Repeat 3 times.

4. PEEP Membrane Function

Cleaning and Sterilization/ Disinfection

Function Test Before Use

Always disassemble 
before cleaning.

Test using steps as shown 
in 

Function Test Before Use

 

section.

Reassemble as 
shown in 

Assembly.

Visually inspect each part for damage and cleanliness.  
If not clean repeat cleaning and sterilization process. 

       Caution

Removed damaged, crazed, cracked or worn parts from use.

Rip Stoppers

PEEP membrane inspection:   

Hold membrane between fingers and  
squeeze softly. Check that slit opens fully,  
but not beyond Rip Stoppers. 

       Caution

If any of the above tests fail, disassemble Upright with PEEP, inspect the components, reassemble and repeat the complete procedure in “Function Test Before Use” section. Remove damaged parts from use. 

Patient Port 
Connector

“LAERDAL”

Patient Port Connector

Lip Valve  Membrane

Ventilation Bag

PEEP Cap

PEEP Membrane

Inlet Housing

Inlet Valve

Disk Membrane

Pressure 

Release Valve

Attach PEEP membrane 
(5.a) onto the patient port 
connector before attaching the 
PEEP cap (5.b) during assembly.

Check that PEEP membrane slit 
opens and seals fully. 

Ensure that the ventilation 
bag’s lip is not twisted or 
tucked under in the reassembly 
procedure.

Patient Valve = Patient Port Con PEEP Me PEEP Cap + Lip Valve Membrane 
Inlet Valve = Inlet Valve Disk Me Inlet H Pressure Release Valve

Remove parts using 
aseptic technique.

Wash all parts in a clean tray with clean water and 
dish soap for at least 2 minutes.
Use a scrub or brush to remove any soil.

Rinse parts in clean water to remove all 
soil and dish soap.
Repeat steps until parts are clean.

Before sterilization, 
dry with clean gauze.

1. Disassemble 

2. Manual Cleaning

3. Rinsing

4. Post-Treatment

Pre-Sterilization/ Disinfection

1. Choose one of the following methods:

2. Post-Treatment

Sterilization/ Disinfection

2. Reassemble

1. Inspect

Post-Sterilization/ Disinfection

3.  Function Test

Newborn Mask

Size 0 and 1

       Caution

For proper function, make sure to only use parts marked “LAERDAL” .

Ventilation Bag and Inlet Valve

“LAERDAL”

 Assembly

To reassemble, follow steps 1-6.

Insert

Attach

Attach

Attach

Screw on

Screw on

Screw on

Allow parts to cool. Dry each part 
with sterile gauze or air dry in a 
protected space.

Autoclaving

Boil

Boil all parts in clean 
water for 10 min. 
Validated at 
approximately
sea-level pressure

Attach

Sterilize by gravity 
displacement steam
autoclaving at 136 °C 
and 2.0 kg/cm

2

 for

10-20 min.

OR