Steam 136 °C
10-20 minutes
100 °C
10-20 minutes
5.b
5.a
3
Disassembly
Overview
Items Included
Assembly Highlights
Upright with PEEP
Newborn Bag-Mask
Reprocessing Instructions
Pull Out
Pull Out
Pull Off
Lift Out
Pull firmly on tab to release
To disassemble follow steps 1-6.
Unscrew
Unscrew
Unscrew
Caution
Do not disassemble parts beyond the steps shown.
1. Lip Valve Membrane Function
2. Pressure Release Valve & Inlet Valve Function
3. Product Sealing
Squeeze the bag. Check that the yellow valve opens
and closes with every squeeze.
Seal the mask with a hand. Squeeze the bag forcefully.
Check that air is released from the pressure release valve.
Keep the mask sealed against hand. Release the squeezed
bag. Check that the bag re-expands without resistance.
Keep the mask sealed against the hand.
Press the pressure release valve down.
Squeeze the bag and check that there is no
leakage.
Hold Upright firmly at the neck. Press other hand
upwards against the mask’s sealing cuff so that air
inside the mask is pushed into Upright and out the
expiration port with the PEEP Valve membrane.
Repeat 3 times.
4. PEEP Membrane Function
Cleaning and Sterilization/ Disinfection
Function Test Before Use
Always disassemble
before cleaning.
Test using steps as shown
in
Function Test Before Use
section.
Reassemble as
shown in
Assembly.
Visually inspect each part for damage and cleanliness.
If not clean repeat cleaning and sterilization process.
Caution
Removed damaged, crazed, cracked or worn parts from use.
Rip Stoppers
PEEP membrane inspection:
Hold membrane between fingers and
squeeze softly. Check that slit opens fully,
but not beyond Rip Stoppers.
Caution
If any of the above tests fail, disassemble Upright with PEEP, inspect the components, reassemble and repeat the complete procedure in “Function Test Before Use” section. Remove damaged parts from use.
Patient Port
Connector
“LAERDAL”
Patient Port Connector
Lip Valve Membrane
Ventilation Bag
PEEP Cap
PEEP Membrane
Inlet Housing
Inlet Valve
Disk Membrane
Pressure
Release Valve
Attach PEEP membrane
(5.a) onto the patient port
connector before attaching the
PEEP cap (5.b) during assembly.
Check that PEEP membrane slit
opens and seals fully.
Ensure that the ventilation
bag’s lip is not twisted or
tucked under in the reassembly
procedure.
Patient Valve = Patient Port Con PEEP Me PEEP Cap + Lip Valve Membrane
Inlet Valve = Inlet Valve Disk Me Inlet H Pressure Release Valve
Remove parts using
aseptic technique.
Wash all parts in a clean tray with clean water and
dish soap for at least 2 minutes.
Use a scrub or brush to remove any soil.
Rinse parts in clean water to remove all
soil and dish soap.
Repeat steps until parts are clean.
Before sterilization,
dry with clean gauze.
1. Disassemble
2. Manual Cleaning
3. Rinsing
4. Post-Treatment
Pre-Sterilization/ Disinfection
1. Choose one of the following methods:
2. Post-Treatment
Sterilization/ Disinfection
2. Reassemble
1. Inspect
Post-Sterilization/ Disinfection
3. Function Test
Newborn Mask
Size 0 and 1
Caution
For proper function, make sure to only use parts marked “LAERDAL” .
Ventilation Bag and Inlet Valve
“LAERDAL”
Assembly
To reassemble, follow steps 1-6.
Insert
Attach
Attach
Attach
Screw on
Screw on
Screw on
Allow parts to cool. Dry each part
with sterile gauze or air dry in a
protected space.
Autoclaving
Boil
Boil all parts in clean
water for 10 min.
Validated at
approximately
sea-level pressure
Attach
Sterilize by gravity
displacement steam
autoclaving at 136 °C
and 2.0 kg/cm
2
for
10-20 min.
OR