User Guide

Upright 

with PEEP

Newborn Bag-Mask

REUSABLE - AUTOCLAVABLE

SPECIFICATIONS

Temperature and Environment

Operating temperature

- 18 °C to 50 °C

Storage temperature

- 40 °C to 60 °C

Performance Specifications

Dead space

Approx. 5 mL (water volume)

Tidal volume

>150 mL

Total bag volume

Approx. 320 mL

Expiratory resistance

with PEEP: <15 cmH

2

O at 5 L/min

without PEEP: <2.5 cmH

2

O at 5 L/min

Inspiratory resistance

with PEEP: <1 cmH

2

O at 5 L/min

without PEEP: <1 cmH

2

O at 5 L/min

PEEP value/Pressure range

6 ± 2 cmH

2

O

Lifetime parameters

Shelf-life

5 years

Expected Service Life

100 cycles of reprocessing

Measurements

Patient Connector 
(conical)

15 mm inner diameter,
22 mm outer diameter

External dimensions
(with newborn Mask size 1)

Approximately 74 x 87 
x 217 mm

Mass 
(with newborn Mask size 1)

Approximately 200 
grams

Materials

Hard plastic components

Polysulfone (PSU)

Soft plastic components

Silicone rubber

Spring

Stainless steel

CLINICAL INDICATIONS

Device Description 

Upright with PEEP is a self-inflating manual resuscitator that is intended 
for patients requiring total or intermittent ventilatory support.  

Indication for Use 

Upright with PEEP is intended for patients requiring total or intermittent 
ventilatory support. Ventilation is possible with or without supplemental 
oxygen. 

Intended Use 

Upright with PEEP provides positive pressure ventilation and allows 
spontaneous breathing with a face mask or an artificial airway. 

Intended for patients up to 10 kg (22 lbs). 

Intended Users 

Upright with PEEP is intended to be used by healthcare professionals 
trained in delivering ventilatory support and in the use of manual 
resuscitators. 

Clinical Benefits 

Positive impact on clinical outcome, by respiratory support that reduces 
probability of adverse outcomes, such as morbidity and mortality caused 
by hypoxia. 

Clinical Outcome 

Desired outcome of ventilation is oxygenation of the patient, often 
evaluated using SpO

2

, EtCO

2

, blood gas analysis or other method of 

analysis. 

Known Side Effects 

Gastric Insufflation 
Oxygen Toxicity 

Contraindications 

No known contraindications for use.

IMPORTANT INFORMATION

Read this User Guide and become familiar with the operation and 
maintenance of the device prior to use. Use the product only as 
described in this User Guide.

  Warnings and Cautions

A Warning states a condition, hazard, or unsafe practice that
can result in serious personal injury or death.
A Caution states a condition, hazard, or unsafe practice that
can result in minor personal injury or damage to the product.

   Notes

A note states important information about the device or its operation.

 

 Warnings

• 

Upright with PEEP should only be used by persons who received 
sufficient training in its use. Incorrect operation of the resuscitator can be 
hazardous. 

• 

Resuscitators should not be used with supplemental oxygen where 
smoking is permitted or when fire, flame, oil or grease is in close 
proximity.  

• 

This resuscitator should not be used in toxic or hazardous atmosphere. 

• 

Do not use Upright with PEEP if you have any reason to be concerned 
about its functionality. 

• 

For proper function, ensure that Upright with PEEP components are 
not mixed and confused with similar-looking non-Laerdal components. 
All Upright with PEEP components are marked LAERDAL. as shown on 
page 2. 

• 

Care should be taken when using the Upright with PEEP on patients 
with severe pulmonary disease or severely immature lungs. Applied 
pressure should be adjusted and monitored according to the patient’s 
condition. Note that a manometer is not supplied by Laerdal for use 
with the Upright with PEEP, but a manometer is possible to connect to 
the patient port with an appropriate adapter compatible with an ISO 
5356-1 connector. 

• 

Care should be taken when using Upright with PEEP on patients with 
severe anomalies or when applying other medical devices which may 
conflict with the mask as mask leakage may occur. If mask face sealing 
is not possible to achieve consider using alternative airway device.  

• 

Care should be taken when applying pressure to the mask to avoid 
facial damage.  

• 

Use of the Newborn Masks provided on patients over 5 kg may result in 
poor fit. Larger masks should be used on patients over 5 kg. Note that 
these masks are not provided with Upright with PEEP.  

REGULATORY

Meets ISO 10651-4:2002/EN ISO 10651-4:2009, Lung ventilators – 
Particular requirements for operator–powered resuscitators.

• 

Care should be taken when using Upright with PEEP on patients 
with severely congested airways. Consider removing congestion from 
the oropharyngeal airway. Use of Upright with PEEP on patients with 
severely congested airways may result in a reduction in expected 
oxygenation. 

• 

Upright with PEEP is not intended for use in delivery of medications, 
such as anesthetic gases.  

• 

Upright with PEEP is not intended for use with advanced airways. 

• 

Do not pull open slit on PEEP membrane. 

• 

Do not attempt to block the PEEP cap or port, which could prevent 
patient expiration. 

• 

High ventilation rates (> 30 breaths per minute) with high-compliance 
lungs, may restrict patient expirations and increase PEEP above 8 
cmH

2

O. 

• 

Use of Newborn PEEP on infants larger than 5 kg may result in higher 
PEEP. 

• 

At freezing temperatures, PEEP function could be reduced or lost. 
Absence of a hissing sound after each ventilation may indicate this.  

• 

Not to be used in a respiratory circuit or for providing CPAP. 

• 

Not suitable for providing free-flow of oxygen to patient.  

• 

Check PEEP levels regularly with a manometer.  

 Cautions

• 

The resuscitator is not intended for use in an ambulance. 

• 

The hard plastic components of the resuscitator and the mask cover are 
incompatible with polar solvents such as ethanol and isopropyl alcohol.  

• 

An oxygen blender is recommended if more precise oxygen 
concentrations are required, for example for pre-terms. 

• 

The use of the PEEP valve is recommended in the case that PEEP is 
indicated for the patient.  

• 

Improper assembly of Upright with PEEP after reprocessing may affect 
performance. 

  Note

• 

Should any serious malfunction, undesirable incident with, or deterioration 
in the functionality or performance of the device occur, contact Laerdal 
promptly. The competent authority where the incident took place and/or 
the device was used should also be notified.  

USING UPRIGHT WITH PEEP

Orientation:

 Upright with PEEP is operated as a normal resuscitator, 

with the bag having a vertical stance over the mask.  

Newborn Mask - Snap Design:

 The mask fits with standard 15 mm 

inner-diameter conical connectors, as defined by ISO 5356-1. Check fit 
before use with other devices. When used with Upright with PEEP, the 
mask attaches with a snap fit when pressed completely into place.

•  Upright with PEEP can provide supplemental oxygen only when used 

with the Oxygen Reservoir Accessory (sold separately).  

•  Upright with PEEP may be reused provided reprocessing procedures 

(page 2) are followed between each patient use. It must be cleaned 
and disinfected before first use.  

•  Pressure Release Valve: Upright with PEEP has a pressure release 

valve (“pop-off ”) which releases air when pressure to the patient 
exceeds 30-45 cm H

2

O. A hissing sound can be heard when the 

valve opens. If higher airway pressure is needed, press downwards on 
the valve with the index finger while squeezing the bag. 

CLINICAL USE

Operating Upright with PEEP with face mask: 
1.  Connect a suitable face mask 

2.  Connect to external O

2

 source, if applicable. 

3.  Place mask over face ard check for seal. 

4.  Squeeze the Ventilation Bag in accordance to clinical protocol. 

5.  Observe patient chest rise during ventilation. 

6.  Allow patient to exhale. 

7.  Stop ventilation as required by clinical protocol.

 

 Warning

Incorrect operation of the resuscitator can be hazardous. 

  Notes

The oxygen source should be able to be adjusted to provide a flow relevant 

to the Upright with PEEP. See tables in Upright Oxygen kit’s user guide 
regarding achievable oxygen concentration at varying flows for more 
information. 

Manometer Use

Peak Inspiratory Pressure (PIP) and PEEP can be monitored when a 
manometer is connected between Upright with PEEP’s Patient Port 
Connector and the face mask. Check PEEP pressure regularly with a 
manometer connected with a manometer connector between Upright 
with PEEP and the mask or airway adjunct. 

Positive End-Expiratory Pressure (PEEP) Valve 

The included Newborn PEEP valve membrane can be attached to 
Upright with PEEP’s expiration port. See section Disassembly and 
Reassembly illustration to locate the expiration port.  

The PEEP membrane opens with a popping sound during each patient 
expiration. Other types of PEEP valves cannot be fitted to Upright with 
PEEP.  

PEEP can be deactivated by removing the PEEP valve membrane. Check 
PEEP pressure regularly with a manometer. 

Contamination: 

If the Patient Valve becomes contaminated with vomitus 

or mucus, disconnect Upright with PEEP from the patient and clear the 
valve as follows: 

•  Tap the Patient Valve with the patient port against your gloved hand 

to shake free any contaminant and squeeze the silicone bag to 
deliver several sharp breaths through the Patient Valve to expel the 
contaminant.  

•  If contaminant does not clear, disassemble the Patient Valve and rinse. 

Reassemble the device and test in accordance with page 2.  

If any components are loose, tighten or reassemble the device in 
accordance with page 2.

Patient port on Newborn 
Mask Snap design fit standard 
15/22 mm connector

Ø 15 mm

Expected PEEP Performance

PEEP shall be measured as the end expiratory pressure between 
inflations (usually the lowest pressure delivered between inflations). 

PEEP value between continuous ventilations is typically 6 cmH

2

O with 

a tolerance range of 4-8 cmH

2

O, as illustrated in the graphs below. For 

high ventilation rates with a high compliance lung, expected PEEP value 
increases.

20

15

10

5

0

Time 

Pr

essur

e  [cmH

2

O]

PEEP

0-50 cmH

2

O

Manometer
Connector

Manometer

20

15

10

5

0

Time 

Pr

essur

e  [cmH

2

O]

PEEP

ACCESSORIES AND SPARE PARTS

Cat. no

Description

846156 
846157 

Newborn Mask - Snap design – Size 0*
Newborn Mask - Snap design – Size 1*

846165

Newborn PEEP membrane and cap – 1 set

846151 

Upright Oxygen kit Complete 

846131

Oxygen reservoir bag and tubing

846155

Upright valves and membranes kit*
(lip valve, inlet valve disc membrane)

* Masks are bulk packed: 10 masks in 1 polybag.

Upright Oxygen kit (Complete)

20

15

10

5

0

Time 

Pr

essur

e  [cmH

2

O]

PEEP

Symbol Glossary

Medical Device

This medical device complies with the general safety and 
performance requirements of Regulation (EU) 2017/745 
for medical devices.

Not made with natural rubber latex

Warranty

Refer to one-year Laerdal Global Warranty for terms and conditions. 
For more information, visit www.laerdal.com.