Temperature Calibration 

   

 

 

 

 

 

 

     

     

 

To  calibrate  the  temperature,  take  a  certified  calibrated  thermometer  in  a  small  bottle  with 
glycerin and place that in the geometrical center of the incubator. 

 

11.1 Low Temperature Point Calibration (0.0) 

a)  Change the set point to lower temperature value, 37°C for HEATING ONLY UNITS (or 8°C for 

REFRIGERATED UNITS,) and let the incubator run for at least 1 hour

– until the temperature is 

constant, and let the temperature inside of the chamber uniform. 

b)  Read the temperature on the thermometer through the glass lid; calculate the difference with 

actual displayed temperature. for example, if reading is 35°C, difference would be 35-37= -2°C, 
while if reading is 38°C, the difference would be 38-37=1°C 

c)  Press  the  "Set/Confirm"  button  and  go  with  the  up  arrow 

to code “7”. Press  the  Set/Confirm 

again,  to  enter  the  Step  1.  "Low  Temperature  Point  Correction  (0.0)",  the  display  shows  the 
"Temperature  Calibration"  with  "0.0"  and  the  "Current  Calibration  Value".  Use  the  up  and/or 
down  key  to  make  a  further  adjustment  on  the  current  correction  value  by  the  temperature 
difference calculated above, for "-2°C ", decrease by 2, for "1°C ", increase by 1. 

d)  Keep  pressing  "Set/Confirm"  button,  skip  the  "  High  Temperature  Point  Correction  (100.0)" 

setting, save and exit. 

e)  The  displayed  temperature  should  have  changed  due  to  the  calibration.  Waiting  for  another 

one hour to let the temperature stabilize again at 37°C, and check the thermometer value again, 
and calculate the new difference.   

f)  if necessary, perform the calibration again until the actual display value equals to the calibrated 

thermometer value. Thus the Step 1. "Low Temperature Point Correction (0.0)" is completed. 
 

11.2 High Temperature Point Calibration (100.0) 

a)  If the incubator is to be used for more than one temperature setting, and "High  Temperature 

Point Calibration (100.0)" needs to be performed as well.   

b)  Change the set point to a higher temperature point as required, for example 60.0°C and let the 

incubator  run  for  at  least  1  hour

–  until  the  temperature  is  constant,  and  let  the  temperature 

inside of the chamber is uniformed 

c)  Change  the  enter  the  code  "7"  again  and  press  "SET"  skip  the  "0.0",  and  enter  the  "100.0" 

page. 

d)  Repeat the same procedure, like the "Low Temperature Point Calibration", to correct the high 

temperature  point,  until  the  display  value  equals  to  the  actual  thermometer  value.  Then  the 
"Step 2 High Temperature Point Calibration (100.0)" is completed. 
 

 
 
 
 
 
 
 
 
 
 
 

          

   

Special Attention! 
 

Temperature  of  each  unit  has  been  carefully  calibrated  in  factory 
before dispatch. 
 
No  further  calibration  is  needed.  But  if  it  does,  do  follow  the 
calibration  process  strictly  or  consult  your  supplier,  as  wrong 
operation can interfere the temperature accuracy significantly. 
 

0.0 Point 

Calibration 

Exit 

100.0 Point 

Calibration 

Summary of Contents for ZWY-111B

Page 1: ...User s Manual This Manual applies to Modes ZWY 111B ZWY 211B ZWY 111G ZWY 211G ZWY 111C ZWY 211C ZWY 111D ZWY 211D ZWF 111 and ZWF 211 Economic Horizontal Shaking Incubator...

Page 2: ...agriculture It is widely used for cell culturing hybridization cell aeration and solubility studies Etc Since the date of your purchase of this product after sale service will always be close to you t...

Page 3: ...r is needed 6 A surge protector is recommended to avoid power related faults 7 In case of malfunction or burning smell the unit must be unplugged immediately Use a circuit breaker to cut off the power...

Page 4: ...orm is still moving do not open the lid too much or too long as this might affect the temperature inside 8 The incubator must be kept away from volatile flammable explosive liquids or gases 9 Please k...

Page 5: ...__ 10 9 Temperature Speed Alarm Settings _____________________________ 11 10 Refrigeration Defrost Settings ________________________________ 12 11 Temperature Calibration______________________________...

Page 6: ...r in refrigerated units only has an independent overload protection CFC free refrigerant which causes no damage to ozone layer Applies on ZWY 2 ZWF 2 models Large doublefold tempered glass window and...

Page 7: ...64 82 82 66 100 ml 64 82 82 66 250 ml 40 45 52 45 500 ml 28 33 38 32 750 ml 23 24 26 23 1000 ml 15 15 22 15 2000 ml 11 12 11 3000 ml 8 11 11 5000 ml 6 6 6 Test tube 18ml 208 208 208 208 Inner dimensio...

Page 8: ...n Hold this button for 2 seconds to turn the unit on off the standby mode Temperature button Press this button to display the actual temperature value and the set temperature in turns Speed button Pre...

Page 9: ...amber is deviated over the preset alarm limit the indicator flashes and the alarm beeps Heating indicator When the heater is connected with power supply this indicator will be light This indicator wil...

Page 10: ...eep unscrewing screwing the adjusting bolt on the front left corner with the spanner until the unit is elevated leveled and sits firmly on the four blocks 4 2 The shaking mechanism of all reciprocal s...

Page 11: ...the current status of temperature retaining Keep Temp Mode function press Increase or Decrease to change the setting 6 2 Press Set confirm to exit to the home page 1 Display current control mode and t...

Page 12: ...imer page press Increase or Decrease to change the value if necessary If Timer is set as 0 the unit will keep running continuously 7 2 Press Set confirm to exit to the home page Special Attention 1 Th...

Page 13: ...ess Increase or Decrease to change the time press Set confirm to save and switch to year setting page press Increase or Decrease to set the year and press Set confirm again to save and switch to the m...

Page 14: ...Increase or Decrease to change the value 9 3 Press Set confirm to exit to the home page Special Attention 1 Deviation alarms can only be activated when unit is running and once actual temperature or...

Page 15: ...ss the Set Confirm button to confirm and store the changed value 10 5 Explanation for setting auto defrosting cycle a Defrost Timer is the time interval between two cycles For example if this is set a...

Page 16: ...it will start manual defrost for set defrosting period To abort press button and hold for 3 seconds Minimum time interval between two manual defrost cycles is 1 min Attention 1 The defrosting cycle wi...

Page 17: ...ture should have changed due to the calibration Waiting for another one hour to let the temperature stabilize again at 37 C and check the thermometer value again and calculate the new difference f if...

Page 18: ...tton if the symbol A is displayed on the lower left corner of the screen it means the non volatile memory function has been active 13 Setting of communication address for RS485 connection Opt If the R...

Page 19: ...on within 3 seconds the remaining operating time will be cleared to zero Press the Start Stop button once more and the unit starts again to count down from the preset operation time 15 5 While the uni...

Page 20: ...through the opening 16 5 Use the Allen key to loosen the A central fixing nut 16 6 Use the 8 10 spanner to hold the B adjusting bolt as still in the meantime loosen the C safety nut with 17 19 spanne...

Page 21: ...cold air is excessive Close a part of the ventilation opening The ventilation fan is broken Notify the distributor to replace the fan The heater does not work Notify the distributor to repair the heat...

Page 22: ...he equipment if it does not work Notify the distributor The equipment causes a strange loud noise The equipment is not placed horizontally Adjust the left back foot to make the equipment stable The fi...

Page 23: ...20 18 Electronic Control System...

Page 24: ...e the parts free of charge Your distributor will provide professional workmanship and will repair the unit without any charge When this one year warranty is expired we will continue to provide you wit...

Reviews: