Maintenance and Service
NXT Go Owner’s Manual, LBL-00145, MAN2030
85
d)
Press the
Zero Weight
button.
e)
The
Relative Weight
for Urocap NXT should measure at 0.0g +/- 2g immediately
following zeroing. To verify calibration, check the absolute and relative weight with
water.
f)
The
Absolute Weight
should measure 0.0g +/- 1kg.
5.3.1.2
Checking the Relative Weight with Water
To complete calibration verification for the Urocap NXT, check the relative weight with
water:
a)
Take a beaker and place it on the Urocap NXT. Wait 10s. Press the
Zero Weight
button.
b)
Add 500 ml of water to the beaker.
c)
Verify that the
Relative Weight
for the Urocap NXT in the diagnostic window is
500.00 g +/-3% as exemplified in Figure 67.
Figure 67: Urocap NXT – Device Diagnostics with Water
Roam NXT
– Verifying Calibration
To setup verification of calibration for the Roam NXT:
1.
Login to the Synergy software.
2.
Ensure the PIM NXT is connected to the Roam NXT. Ensure to connect the PIM NXT
that corresponds with the consumable type in use.
3.
Connect the Urocap NXT, Roam NXT, and the Infusion Pump NXT to the Synergy
software and setup the hardware channels.