GUIDANCE TABLES (continued)
Prima Mµ is intended for use in the electromagnetic environment specified below. The customer or
user of the Prima Mµ should ensure that it is used in such an environment.
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
All Equipment and Systems
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Immunity Test
Compliance Level
ESD
IEC 61000-4-2
± 6kv contact
± 8kv Air
± 2kv Mains
± 1kv I/Os
± 1kv Differential
± 2kv Common
± 1kv Differential
± 2kv Common
± 2kv Mains
± 1kv I/Os
± 6kv contact
± 8kv Air
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
dropouts,
IEC 61000-4-11
Power Frequency
50/60Hz Magnetic
Field IEC 61000-4-8
3A/m
3A/m
IEC 60601 Test
Level
Electromagnetic Environment
- Guidance
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
>95% Dip
for 0.5 Cycle
60% Dip
for 5 Cycles
30% Dip
for 25 Cycles
>95% Dip
for 5 Seconds
Floor should be wood, concrete or ceramic tile. If
floors are synthetic, the R/H should be at least 30%.
Mains power quality should be that of a typical
Commercial or hospital environment.
Mains power quality should be that of a typical
Commercial or hospital environment.
Power frequency magnetic fields should be that
of a typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Prima Mµ requires continued
operation during power mains interruptions,
it is recommended that the Prima Mµ be powered
from an uninterpretable power supply
or battery.