background image

 

 

 

 

  BA_085_2019-05_10.00  Temperature transmitter MiniTherm 

Page 15/20 

 

6.4.4.2 

Submenue "Screen mode" 

In the "Screen mode" submenue, you configure the representation of the measured values 
and additional information on the display. 
With the menue item “Screen layout” you configure the information that is displayed and its 
layout. Up to four values can be displayed at the same time. In the additional menue en-
tries "1st value" to "4th value" you define the contents of the placeholders in the layout. 

The various layouts as well as the possible content of the values are described in Chapter 
6.2.4. 

  

6.4.5 

Menue "Measurement/Output" 

Configure the measured value as well as the current output  in the "Measurement/output" 
menue. 

  

Menue entry

 

Description

 

Output function 

Setting the output function (linear, inverse) 

Lower range val. 

Setting the temperature value that shall correspond to 4 mA (start 
of range) 

Upper range val. 

Setting  the  temperature  value  that  shall  correspond  to  20  mA 
(end of range) 

Alarm current 

Select the alarm current: High (>21 mA) or low (<3.6 mA) 

Minimum current 

Limit of the lower output current (3.8…4.0 mA) 

Maximum current 

Limit of the upper output current (20…21 mA) 

Table 11: "Measurement/Output" menue 

  

Independent from the setting of the minimum and maximum current limit, the set measur-
ing range always corresponds to a current range of 4…20 mA. 

  

6.4.5.1 

Output function 

You can set the output function proportionally rising to the measured value (selection "lin-
ear") or proportionally falling (selection "inverse").  

  

6.4.5.2 

Setting the measurement range 

You can freely set the lower and upper range value within the nominal range of the device. 
Please note that when the turndown is increased, the measurement error also increases. If 
you  use  a  very  small  range  you  must  therefore  accept  a  significant  measurement  error. 
We recommended a maximum turndown of 10:1. 

The lower range value must be below the upper range value. For an inverse characteristic 
curve, select the output function setting “inverse”. 

  

6.4.5.3 

Minimum and maximum current 

In the standard setting, the current output is limited at 3.8 and 20.5 mA, meaning a further 
drop or rise in the measured  variable does not change the current. You can freely select 
these  current  limits  for  the  lower  limit  between  3.8  and  4  mA  and  for  the  upper  limit  be-
tween 20 and 21 mA. 

  

Summary of Contents for MiniTherm GV4730 Series

Page 1: ...afety Notes 2 1 2 Intended Use 2 1 3 Conformity with EU Regulations 2 2 Transportation and Storage 2 3 Installation and Commissioning 2 3 1 Mechanical Installation 2 3 2 Mount Dismount Display 3 3 3 E...

Page 2: ...data sheet 1 3 Conformity with EU Regulations The CE marking on the device certifies its compliance with the applicable EU Directives for placing products on the market within the European Union You f...

Page 3: ...0 mA 20 4 mA Permissible supply voltage UV 12 30 VDC Permissible load RB UV 12 V 22 mA Circular connector M12 Cable gland Figure 1 Options for the electrical connection Pay attention to the following...

Page 4: ...play module Cable glands if applicable For defects to other components repairs must be performed in the factory 5 Disassembly When measuring hot media make sure that the device has cooled down prior t...

Page 5: ...on of the device The gen eral functionality of the buttons is identical in all operating modes Figure 2 Control elements If you press and hold the up or down button it will automatically be triggered...

Page 6: ...value is larger than the displayable number due to setting a fixed decimal point or an inappropriate unit the largest displayable number is shown flashing In this case please select a different setti...

Page 7: ...text Write protection in combination with the lock icon Use the same key combination to unlock the menue again 6 2 3 Icons indicating device status In the top right corner of the display depending on...

Page 8: ...ain value is shown at the maxi mum size best readability from a dis tance One additional value can also be shown Table 4 Display layouts You can define what information shall be shown in the layout s...

Page 9: ...by pressing and holding the ESC button for at least one second Figure 4 Navigating the operating menue In long menues and selection lists a scroll bar on the left side shows the position of the curren...

Page 10: ...ist is longer a scroll bar on the left side of the display indicates the position of the selected item in the selection list When you press the ESC button you cancel the entry and go back to view mode...

Page 11: ...e selected digit by pressing the or button The higher value digit is also increased or decreased when passing zero For instance you can easily go from 100 to 90 without having to edit two digits Lower...

Page 12: ...rature and current for function check of the measurement chain Communication Information and settings regarding HART communication System Device data and function such as factory reset Tabelle 6 The m...

Page 13: ...t for a correct full adjustment Figure 8 Effect of the upper and lower adjustment on the characteristic curve You can perform the upper and lower adjustment at any reference temperature level For inst...

Page 14: ...In the Display menue you find all the settings that affect the display on the screen Menue entry Description Language Select menue language Units Setting units for the different measured and displayed...

Page 15: ...Limit of the lower output current 3 8 4 0 mA Maximum current Limit of the upper output current 20 21 mA Table 11 Measurement Output menue Independent from the setting of the minimum and maximum curren...

Page 16: ...ble from the measured value display in the device data see 6 2 1 Figure 9 Display and resetting of the Min Max values 6 4 7 Menue Simulation In the Simulation menue you can simulate the temperature as...

Page 17: ...ice ID e g to display a free text in the display Device data Display of device data such as from measured value display see 6 2 1 Factory reset Reset to factory settings Device reset Restarting the de...

Page 18: ...BA_085_2019 05_10 00 Temperature transmitter MiniTherm Page 18 20...

Page 19: ...BA_085_2019 05_10 00 Temperature transmitter MiniTherm Page 19 20...

Page 20: ...BA_085_2019 05_10 00 Temperature transmitter MiniTherm Page 20 20 6 4 10 Overview with menue tree and device functions...

Reviews: