BA_084_2019-05_10.00 Temperature transmitter clamp-on
Page 9/20
The sequence of the screens with device data is as follows:
Min/Max values (measurement)
----- Measured value display (starting point) ---
Temperature measurement (senor type)
Current output (characteristic curve, limits, measuring range)
HART data (address, tag, descriptor)
Device identification (device ID, order number, serial number)
Module information (hardware and software versions, serial numbers)
6.2.2
Locking the Menue
You can lock the menue with a key combination at the device. Press and hold the ESC-
key and press then OK to activate the menue lock. Hold down both keys until the display
confirms that the menue lock has been activated. You can still access the device data but
the menue is now locked.
An active menue lock is indicated by the header text „Write protection“ in combination with
the lock icon.
Use the same key combination to unlock the menue again.
6.2.3
Icons indicating device status
In the top right corner of the display, depending on the device status an NE107-conform
icon is displayed. The following icons are defined:
Error/failure
Critical error, alarm current is activated.
The error description is indicated in the display. The device
restarts after confirming the error.
Warning
Temperature limit is exceeded.
Limit reached
The output current reached the pre-set upper or lower cur-
rent limit.
Function control The output current does not correspond to the measured
value due to an active simulation.
Write protection The write protection of the device was activated via HART
or with the key combination in the display.
Table 3: Icons for device state
The list of icons is sorted in descending priority. Only the icon with the highest priority is