Mars-Mars PRO
User Manual
Code No. 9.900.900.710
U:\QA\Manualer\Arkiv\Manuals- Labogene-Origio-GKO\Manualer-WORD\ScanLaf\Mars Mars Pro manual-UK - rev. 06.doc
Page 26 af 34
1. Dismantle the exhaust system, and close the exhaust opening tight with a sealed plate
2. Disconnect the exhaust fan from by removing the fuses (fan 3).
3. Prepare the formaldehyde- and the neutralization solution and place in the evaporator and place it
inside the Mars-Mars pro. (The evaporator has a timer function).
4. Bring the sliding window to the lowest position and open the front cover.
5. Seal the window by means of adhesive tape.
Procedure for decontamination in the Mars-Mars Pro
1 Let the Mars-Mars Pro run with reduced speed during evaporation of the formaldehyde. This
distributes the formaldehyde in the Mars-Mars Pro.
2 Let the influence with formaldehyde last for 6 hours, and let the neutralization last for ca. 1 hour.
3 Switch off the fan.
4 Dismantle the metal plate on the exhaust opening, and mount the exhaust connection.
5 Connect the exhaust fan, reinstalling the fuses.
6 Remove the tape on the front window.
7 Close the front cover and set the sliding window in working position.
8 Let the Mars-Mars Pro run for at least 30 minutes.
9 The Mars-Mars Pro is ready for use.
Decontamination and possible neutralization agents
(The formaldehyde is diluted with water)
:
Model ScanLaf Mars - Mars Pro 900:
Formaldehyde 36%: 54 ml; Water: 54 ml: Ammonia 25%: 38 ml
Model ScanLaf Mars - Mars Pro 1200:
Formaldehyde 36%: 72 ml; Water: 72 ml: Ammonia 25%: 50 ml
Model ScanLaf Mars - Mars Pro 1500:
Formaldehyde 36%: 90 ml; Water: 90 ml: Ammonia 25%: 63 ml
Model ScanLaf Mars - Mars Pro 1800:
Formaldehyde 36%: 108 ml; Water: 108 ml: Ammonia 25%: 75 ml
ScanLaf refers to our Formaldehyde Vaporizer catalogue number: 9.001.020.117.
Decontamination procedure
Formaldehyde evaporation
30 minutes
Effect time
8 hours
Neutralization
30 minutes
Effect time
30 minutes