6.13
Disposal of cabinets
All components except for the HEPA filter can be discarded after having been thoroughly cleaned
and decontaminated. The HEPA filter must be discarded in accordance with the applicable
international, national and local regulations for chemical or biological waste.
Warning
As this cabinet can be used for processing and treating infectious substances, it must
be decontaminated prior to disposal, in accordance with acceptable standards and
procedures.
7
Cleaning/Decontamination of cabinets
Several procedures can be used for decontaminating the biological safety cabinet. Which procedure
is selected, depends on the potential risk imminent in the products handled, and/or the degree of
cleanliness required by an experiment or work process.
One possible decontamination procedure is to clean, rinse and dry with a disinfection liquid or spray,
and sterile distilled water.
UV disinfection can also be used. It is particularly suited as a more intense disinfection after the
above procedure.
Decontamination using hydrogen peroxide H
2
O
2
can be performed if a sterile working chamber is
required for the work procedure. This decontamination procedure or a similar one is required before
filters are replaced, or before the cabinet is discarded. The control system incorporates a procedure
for this process.
Decontamination using formaldehyde can be performed if a sterile working chamber is required for
the work procedure. This decontamination procedure or a similar one is required before filters are
replaced, or before the cabinet is discarded. The control system incorporates a procedure for this
process.
7.1
Caring for and cleaning of stainless steel
Stainless steel is a combination of many different metals including iron and chromium. Iron, the
primary element in stainless steel, tends to corrode (rust) when in its natural state. The chromium
content in stainless steel prevents corrosion.
Stainless steel is not corrosion or rust proof, just resistant to stains, or
stains “less”. The chromium,
in the presence of oxygen, forms a tough, invisible, passive layer of chromium oxide film on the steel
surface. If damaged mechanically or chemically, this film is self-healing as long as it has enough
oxygen. The presence of any liquid or solid that remains in contact with the stainless steel for a
prolonged time can prevent oxygen contact and promote corrosion, as can prolonged contact with
cleaners or disinfectants containing chlorine, ammonia, iodine or other caustic agents.
1.
To properly care for stainless steel, use cleaners and disinfectants free of caustic agents
such as chlorine, iodine and ammonia.
2.
Always follow the application of any cleaner or disinfectant with a minimum of two clean
distilled water rinses, then a thorough drying with a clean soft cloth. By rinsing with distilled
water and drying, any remaining residue is removed from contact with the stainless steel.
Summary of Contents for Fortuna 900
Page 39: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 39 of 49...
Page 40: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 40 of 49...
Page 41: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 41 of 49...
Page 42: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 42 of 49...
Page 43: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 43 of 49...
Page 44: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 44 of 49...
Page 45: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 45 of 49...
Page 46: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 46 of 49...