Instruction manual for Fortuna
12062017QA01
Item No. 700745
CopyRight © 2017
Page
14
of
49
If a hard-wired connection is used, a main switch, which will cut off all poles to the unit, must be
used. This switch must be lockable both in the ON and OFF positions.
5.3
Assemble the stand for the Fortuna cabinets
As default the Fortuna cabinets are delivered without the stand mounted.
Stand
Electrical stand
All nuts and bolts for assembly of the stand and for mounting the stand to the cabinet are delivered
with the stands.
Start by assembling the stand and then place the cabinet on top and fasten with four bolts
– one in
each corner.
Product name
Catalogue No.
Stand, Fortuna 900, 75-80 cm
9.001.020.011
Stand, Fortuna 1200, 75-80 cm
9.001.020.012
Stand, Fortuna 1500, 75-80 cm
9.001.020.013
Stand, Fortuna 1800, 75-80 cm
9.001.020.014
Stand, Fortuna 900, 80-85 cm
9.001.040.011
Stand, Fortuna 1200, 80-85 cm
9.001.040.012
Stand, Fortuna 1500, 80-85 cm
9.001.040.013
Stand, Fortuna 1800, 75-80 cm
9.001.040.014
Stand, Fortuna 900, 90-95 cm
9.001.030.011
Stand, Fortuna 1200, 90-95cm
9.001.030.012
Stand, Fortuna 1500, 90-95cm
9.001.030.013
Stand, Fortuna 1800, 90-95 cm
9.001.030.014
Electrical stand, Fortuna 900, 75-105 cm
9.001.020.101
Electrical stand, Fortuna 1200, 75-105 cm
9.001.020.102
Electrical stand, Fortuna 1500, 75-105 cm
9.001.020.103
Electrical stand, Fortuna 1800, 75-105 cm
9.001.020.104
5.4
Levelling of the Fortuna cabinets
To ensure the best working conditions possible the cabinet must be levelled off before getting
released for normal operation.
Place a levelling device directly on the table tops and adjust the feet on the stand until the cabinet is
perfectly in horizontal level. Make sure that the cabinet is in horizontal level from side to side and
from front to back.
5.5
Power connection of the Fortuna cabinets
The Fortuna cabinets are connected to power through the delivered power cable. It is mounted in
the power inlet placed on top of the cabinet and to a standard power socket. All that is required is
230 VAC / 50 Hz connected to a 10 A or 13 A fused socket in the general electrical installation.
Summary of Contents for Fortuna 1200
Page 39: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 39 of 49...
Page 40: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 40 of 49...
Page 41: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 41 of 49...
Page 42: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 42 of 49...
Page 43: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 43 of 49...
Page 44: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 44 of 49...
Page 45: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 45 of 49...
Page 46: ...Instruction manual for Fortuna 12062017QA01 Item No 700745 CopyRight 2017 Page 46 of 49...