Page 1: ...Labnet VorTemp 56 Shaking Incubator Instruction Manual Catalog Numbers S2056A S2056A 230V...

Page 2: ...orm 3 5 0 Safety Information 4 6 0 Installation 4 6 1 Device Placement 4 6 2 Attaching the Power Cord 4 6 3 Start Up Check List 4 7 0 Operating Instructions 5 7 1 Control Panel 5 7 2 Basic Operation 5...

Page 3: ...Electrical S2056A S2056A 230V 120V 10 50 60Hz 4 8A 230V 10 50 60Hz 2 4A Fuse S2056A S2056A 230V 2 x T 6 3A L 250V 2 x T 3 15A L 250V Environmental Temperature From 5 C to 40 C Relative Humidity Up to...

Page 4: ...temperature sensor and the control electronics regulate heater function to maintain the set temperature 4 0 Accessories Cat No Description Qty Pk S2056 R 1 5 2 0 mL platform workstation for microcentr...

Page 5: ...stacked The working temperature range is ambient from 5 C to 40 C Please note that the normal maximum temperature for ordinary microplates is below 60 C The platform is made of stainless steel and is...

Page 6: ...ss to the incubator control panel and main plug in case of emergency When selecting the right place for device please consider following Put the device on a smooth horizontal and stable place Leave at...

Page 7: ...ture 6 Encoder Knob Rotate the encoder right or left to modify the Time Temperature or RPM settings Push in the Encoder knob to change between Time Temperature and RPM set values If you rotate the Enc...

Page 8: ...Temp C Start RPM Illuminated Rotating the encoder right or left sets the rotating speed to the desired value 34 340 rpm 90 900 rpm By turning the RPM setting below 20 the shaking function can be turn...

Page 9: ...left until the new value is set While adjusting the RPM the light will pulse When you stop rotating the encoder knob the RPM light will stop blinking after 2 sec l l l l l l l l l 37 0 Stop Encoder Po...

Page 10: ...with the motor regulation PWM regulator pulse generator motor The unit will automatically stop Call for service E 1 2 The motor does not reach the set RPM in 30 sec The unit will automatically stop Ca...

Page 11: ...chase CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be...

Page 12: ...SIA PACIFIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 India t 91 124 4604000 Japan t 81 3 3586 1996 Korea t 82 2 796 9500 Singapore t 65 6572 9740 Taiwan t 886 2 2716 033...