background image

Figure 3.  Loading the rotor

Figure 2.  Bottom of angle rotor

5

3.5  Overloading the rotor

The maximum load of the rotor and the maximum speed have
been established by the manufacturer.  Do not attempt to
exceed these values.
The maximum speed of the rotor has been measured for liquids
having a homogeneous density of 1.2g/ml or less.  In order to
centrifuge liquids with a higher density it is necessary to reduce
the speed.

Failure to reduce the speed may result in damage to the
rotor and centrifuge.

The revised maximum speed can be calculated with the following
formula:

Reduced speed (n

red

) =

x max speed (n

max

)

Example:

Where the density of the liquid is 1.7, the new maximum speed
would be calculated as follows:

n

red 

 =

  x 6,500 = 5,461 rpm

If in doubt concerning maximum speeds, please contact the
manufacturer for assistance.

4.  Operation

ATTENTION: Never attempt to operate the centrifuge with
rotors or adapters that show signs of corrosion or mechanical
damage.  Never centrifuge strongly corrosive materials that
may damage the rotors or accessories.

               1.2
  higher density value

 1.2
 1.7

6

Figure 1.  Chamber and motor shaft

Summary of Contents for Spectrafuge 6C

Page 1: ...rafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Spectrafuge 6C Compact Centrifuge Labnet International PO Box 841 Woodbridge NJ 07095 Phone 732 417 0700 Fax 732 417 1750 Email labnet labnetlink...

Page 2: ...rotor area unless the rotor is completely stopped NEVER move the centrifuge while the rotor is spinning NEVER use solvents or flammables near this or other electrical equipment NEVER centrifuge flamma...

Page 3: ...ed selection Selection of operating time and momentary operation Starting the Centrifuge 5 Service and Maintenance 8 Service Cleaning Disinfection Replacing fuses 6 Troubleshooting 9 7 Where to Call 1...

Page 4: ...properly balanced 1 Possible reason Lid lock is jammed Solution Call for service Possible reason Centrifuge is not receiving power Solution See Centrifuge will not start Centrifuge cannot be started a...

Page 5: ...aning the rotor The rotor should be cleaned after each use When spinning samples containing phenol or phenol chloroform the rotor should be cleaned immediately after use 5 4 Disinfection Should a spil...

Page 6: ...of operating time and momentary operation Time can be selected in half minute intervals from 0 5 to 10 minutes and in one minute intervals from 11 to 30 mintes Time can also be set to continouos hold...

Page 7: ...on the large LED display right Figure 4 Spectrafuge controlpanel layout 7 Tube and tube adapter reference chart 3 2 Rotor maintenance The rotor should be cleaned thoroughly after each use Thor ough cl...

Page 8: ...sult in damage to the rotor and centrifuge The revised maximum speed can be calculated with the following formula Reduced speed nred x max speed nmax Example Where the density of the liquid is 1 7 the...

Reviews: