4.2  Lid release 

Following a run, the centrifuge will display will show flashing “00”.  This 
signals the end of a run and the lid can now be opened by pressing the “lid 
knob” (left).  Note that the   lid can not be opened until the display flashes 
“00” and the rotor has stopped. 

WARNING:  Do not attempt to open the lid of any centrifuge 
until the rotor has come to a complete stop. 

In the event of a power failure or malfunction, it may be 
necessary to open the lid manually. 

1.  Disconnect the power cord from the wall socket. 
2.  Remove the plastic plug, located on the left side of the unit, below the 

quick button. 

3.  Pull the cord (attached to the plug) to open the lid lock manually. 

4.3  Lid lock 

The centrifuge can be started only with the lid securely closed.   Do not 
attempt to   open the lid until the end of run signal “00” is displayed. 

4.4  Speed selection 

 

The speed (rpm) can be selected from 300 to 6,500rpm with the knob (right).   

 

The set speed can be viewed at all times on the large LED display (right). 

 

4.5  Selection of operating time and momentary operation 

Time can be selected in half minute intervals from 0.5 to 10 minutes, and in 
one minute intervals from 11 to 30 minutes.  Time can also be set to 
continuous/hold by turning the timer knob past the thirty minute position.  
This will display the continuous setting “On”. 

When the preselected time expires, the centrifuge will stop automatically.  
To stop the centrifuge prior to the expiration of set time, press the 
“start/stop” knob. 

The centrifuge may be operated manually by pressing and holding the 
“start/stop” knob.  The centrifuge will continue to run as long as the knob is 
depressed. 

4.6  Starting the centrifuge 

 

Once the time and speed have been set the centrifuge can be started by 
pressing the “start/stop” knob.  The centrifuge will then run for the specified 
amount of time. 

 

 

 

 

 

Summary of Contents for Spectrafuge 6C

Page 1: ...Lit M00031 Rev 2 August 2016 SPECTRAFUGE 6C User Manual C0060 C0060 230V...

Page 2: ...is designed to assist you in the optimal usage of your Labnet Spectrafuge 6C Centrifuge The manual is available in English French German Italian Portuguese and Spanish on our website at www labnetinte...

Page 3: ...same type and weight ALWAYS locate the centrifuge within easy access to an electrical outlet ALWAYS use only centrifuge tubes designed to withstand centrifugal forces of at least 4 000 xg ALWAYS use a...

Page 4: ...3 Installing the Rotor 2 Rotors and accessories Rotor maintenance Removing and installing the angle rotor Overloading rotors 4 Operation 5 Closing the lid Lid release Lid lock Speed selection Selecti...

Page 5: ...any of the following conditions exist The centrifuge has not been installed properly The centrifuge is partially dismantled Service has been attempted by unauthorized or unqualified personnel The rot...

Page 6: ...m the carton Retain the carton and cushions until it has been established that the centrifuge is working properly Inspect the centrifuge for any visible signs of shipping damage Shipping damage is the...

Page 7: ...5mm 13x100mm or common Sarstetd style tubes Tube and tube adapter reference chart 3 2 Rotor maintenance The rotor should be cleaned thoroughly after each use Thorough cleaning must be performed when s...

Page 8: ...ng page When loading the rotor refer to figure 3 located on the following page Loading in the pattern indicated will ensure a balanced load Tubes to be loaded should be filled equally by eye The diffe...

Page 9: ...th the following formula Reduced speed nred x max speed nmax Example Where the density of the liquid is 1 7 the new maximum speed would be calculated as follows nred x 6 500 5 461 rpm If in doubt conc...

Page 10: ...00 is displayed 4 4 Speed selection The speed rpm can be selected from 300 to 6 500rpm with the knob right The set speed can be viewed at all times on the large LED display right 4 5 Selection of oper...

Page 11: ...infectious materials occur within the rotor or chamber the unit should be disinfected This should be performed by qualified personnel with proper protective equipment 5 5 Replacing fuses Check the fus...

Page 12: ...ed from 8 30am to 5 30pm EST Monday through Friday Our 24 hour fax number is 732 417 1750 Inquiries may also be sent via our electronic mailbox at labnetinfo corning com Should your Spectrafuge requir...

Page 13: ...9...

Page 14: ...and electronic equipment WEEE Labnet Spectrafuge 6C is marked with the crossed out wheeled bin and must not be disposed of with domestic waste Consequently the buyer shall follow the instructions for...

Page 15: ...aint or finish Claims for transit damage should be filed with the transportation carrier ALL WARRANTIES INCLUDING THE IMPLIED WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMIT...

Page 16: ...9299870000 31 Mayfield Ave Edison NJ 08837 P 732 417 0700 F 732 417 1750 Toll Free 888 522 6381 labnetinfo corning com www labnetinternational com...

Reviews: