9.0 Creating a New Protocol
A new Protocol can be created by going to the User folder and selecting and editing a basic Protocol format, and then assigning it a
new name.
1. Select User name on the User screen. Press F1:OPEN or ENTER.
2. Key in PIN to enter the User’s folder.
3. Press F2:NEW to create a new Protocol by using a standard template, shown below.
NOTE:
Users can create a new Protocol under the Public folder without a PIN. Users can edit a new Protocol inside the Preset folder,
but cannot not save these changes to the original Protocol settings.
Mike>PROTOCOL>NEW
8/21/2022
11:00
1 Tmp
3 Tmp 25 Cyc
2 Tmp 1 Cyc
Temp 94.0
Time 5:00
94.0
55.0
0:30
0:30
72.00
4.00
10:00
SAVE
INSERT
DELETE
ADV
EXIT
1 Tmp.
Segment
3 Tmp. Segments
Tmp./Time
Values
9.1 Changing Temperature/Time Values
1. Press the directional arrows to move the cursor to the temperature or time settings required to be changed.
2. Enter your desired values and press the settings.
The temperature and time ranges are shown below.
Temperature range:
4°C to 99.9°C
Time range (minute:second):
00:01 to 99:59 (99:99 =
∞
)
9.2 Inserting/Deleting Temperature Segment
Temperature Segment can be inserted or deleted by selecting F2:INSERT or F3:DELETE. The inserted or deleted temperature segment
is dependent upon the location on the temperature segment where the cursor is positioned. If the cursor is positioned inside the first
multi-temperature/time segment, pressing F2/F3 will insert/delete the whole Multi-Temperature Cycle. If the cursor is located in any
other Multi-Temperature/Time segment only the selected segment will be inserted/deleted.
9.3 Changing Temperature Segment Count (Temp) and Cycle Count (Cyc)
The Temperature Segment Count (Temp) can be modified by entering a new value (1 to 20) in the first Temperature position and
pressing ENTER. When the Temperature Segment Count is larger than one, the temperature Cycle Count (Cyc) will appear. The
Multi-Temperature Segment will cycle during this segment. The maximum temperature segment per protocol is 12 segments. The
temperature segment count and cycle count ranges are shown below.
Temperature segment count (temp.) range:
1 to 20 cycle
Count (Cyc) range:
1 to 99