Instruction Manual |
1
1.0 Introduction
The Labnet MPS-1000 Mini Plate Spinner is designed for quick spins of samples in PCR plates. Primarily designed for use with 96- or
384-well plates and small volumes, it will accommodate skirted, non-skirted and semi-skirted styles. With the use of the adapter you
may spin half plate, 8 or 12 x 0.2 μL strips, and individual PCR tubes.
No decontamination is required for the equipment.
Before using any cleaning or decontamination methods except those recommended by the manufacturer, check with the
manufacturer that the proposed method will not damage the equipment.
2.0 Symbols and Conventions
The electrical warning indicates the presence of a potential hazard which could result in electrical shock.
CAUTION:
This symbol refers you to important operating and maintenance (servicing) instructions within the product
Instruction Manual. Failure to heed this information may present a risk of damage or injury to persons or equipment.
This symbol identifies a Protective Earth (PE) terminal, which is provided for connection of the supply system’s protective
earth (green or green/yellow) conductor.
3.0 Safety Information
Never use the centrifuge in any manner not specified in these instructions.
Never put hands in the rotor area unless the rotor is completely stopped.
Never move the centrifuge while the rotor is spinning.
Never use solvents or flammables near this or other electrical equipment.
Never centrifuge flammable, explosive or corrosive materials.
Never centrifuge hazardous materials outside of a hood or proper containment facility.
Never continue to operate a centrifuge that is vibrating excessively.
Always load the rotor symmetrically. Each plate should be counterbalanced by another plate of the same brand, type and volume.
Always use plates that have been sealed in some manner to avoid spillage.
Always load plates so that the sealed side faces the center of the rotor and the bottom of the plate faces out.
Always be sure the plates are pushed all the way to the bottom of the rotor slot before operating the centrifuge.
Always locate the centrifuge within easy access to an electrical outlet.
WARNING:
This unit is intended for short run usage only, requiring 15-20 seconds to spin down samples. Inadvertent
continuous use could cause sample temperature to rise and the thermal switch to open, which will shut down the motor.
4.0 Specifications
Capacity
2 PCR plates
Dimensions
7.5 x 8.2 x 7.2 in. (19 x 21 x 18.3 cm)
Electrical
100, 120 or 230V, 50/60 Hz
Max Speed
2,500 rpm ±10%
Max RCF
500 x g
Electrical
100V, 50/60Hz, 0.6A (Fuse 1.0 AT)
120V, 50/60Hz, 0.6A (Fuse 1.0 AT)
230V, 50/60Hz, 0.3A (Fuse 0.5 AT)