Labnet Excel P3600L-10 Instruction Manual Download Page 5

  Instruction Manual        | 

         3

 4.0  Getting  Started

4.1 Unpacking

The package includes:

 



Labnet Excel™ electronic pipettor

 



Battery

 



Battery installation guide

 



AC/DC power adapter

 



Hanger with sticker

 



Pipet tips (sample non-filtered pipet tips: 
1 x 10 µL, 1 x 200 µL, 1 x 1000 µL)

 



Quick start guide

 



Instruction manual

 



Calibration report

If any items are missing, damaged, or incorrect, please contact your distributor or sales representative.

NOTE:

 Charge the battery before first-time operation.

4.2 Inserting the Battery

Remove the battery cover by pressing the battery cover latch and moving the cover outward (Figure 1). Plug in the battery connector 
and insert the battery (Figure 2). Close the cover of the battery compartment. Press the Reset button to power on and calibrate the 
Labnet Excel electronic pipettor. 

NOTE:

 After a charged battery is inserted, the Labnet Excel electronic pipettor will automatically perform a calibration routine.

Figure 2.

 Insert the 

battery connector into 
the bottom right recess 
first, then place the 
battery into position.

Figure 1.

 Open the 

battery cover.

4.3 Charging the Battery 

Attach the included AC/DC power adapter to the pipettor to charge the battery.

Insert the battery into the battery compartment. Insert the DC-in plug of the power adapter into the DC-in jack (Figure 3). Connect the 
power plug into the external power source. 

NOTES:

 



Connecting the power plug before inserting the battery will inactivate the charging process. 

 



Labnet Excel™ electronic pipettors can be used during charging.

 



Do not over-charge the battery for a long period of time.

 



Do not insert or remove the battery during the charging process, as this may cause battery or PCBA damage.

The indicator bar inside the battery symbol (

) will blink during the charging process. When the battery is fully charged (

), 

the blinking will stop and the Labnet Excel electronic pipettor will beep once.

Figure 3.

 Connect the 

AC/DC adapter to 
charge the battery.

Summary of Contents for Excel P3600L-10

Page 1: ...l Single and Multi channel Electronic Pipettors Instruction Manual Catalog Numbers P3600L 10 P3608L 10 P3612L 10 P3600L 20 P3608L 20 P3612L 20 P3600L 200 P3608L 200 P3612L 200 P3600L 1200 P3608L 1200...

Page 2: ...Audible Signals 7 6 4 Blow out 7 6 5 Reverse Pipetting 7 6 6 Program Storing and Recall 7 7 0 AUTO Automatic Pipetting Mode 8 7 1 Setting Up the Volume in AUTO Mode Only 8 7 2 Setting Up the Volume in...

Page 3: ...lly explosive environment or with potentially explosive chemicals When using organic solvents or corrosive chemicals please check their compatibility with the pipet tips and the Labnet electronic pipe...

Page 4: ...5 to 0 5 P3612L 200 LE200 12 10 200 1 3 0 to 0 6 1 0 to 0 15 P3612L 1200 LE1200 12 100 1200 1 3 0 to 0 5 0 8 to 0 2 NOTE Labnet Excel electronic pipettor accuracy and precision specifications have bee...

Page 5: ...pipettor will automatically perform a calibration routine Figure 2 Insert the battery connector into the bottom right recess first then place the battery into position Figure 1 Open the battery cover...

Page 6: ...accuracy and precision specifications for the Labnet Excel electronic pipettors are guaranteed only when using Labnet BioFree tips Contact Corning Customer Service for pipet tip ordering information T...

Page 7: ...mode selection and volume settings This feature can speed up the setting process Enter Button Confirms your choice Count No Prog No Volume L Down Button Decreases volume and changes parameters Up But...

Page 8: ...ion map 6 0 SYS System Mode SYS mode is used to set up the function settings before pipetting Once the user has selected and stored these settings they will no longer need to set up these settings bef...

Page 9: ...tomatic reverse pipetting is to reserve the final drop of sample The Labnet Excel electronic pipettor will not perform reverse pipetting in MIX or SA modes 1 The Reverse Pipetting symbol should now be...

Page 10: ...s the Plunger button to aspirate the sample Press the Plunger button again to dispense the sample and the LCD will display the mixing cycle count and volume Press the Plunger to mix and dispense the s...

Page 11: ...lashes 3 Press the Enter button to confirm the selection 4 Use the Up Down button until the SA symbol flashes 5 Press the Enter button to confirm the selection 6 Under the COUNT symbol 01 will appear...

Page 12: ...y for dispensing 11 The pipettor will make two beeping sounds to remind the user of the last measured volume to dispense At the end of the sequential dispensing cycle if Reverse Pipettor mode is selec...

Page 13: ...on assembly downward until the cylinder lock appears Cylinder lock C Push the metal cylinder lock downward until you hear a click sound D The lower piston assembly will automatically come off easily F...

Page 14: ...the performance of your Labnet Excel electronic pipettor regularly e g every 3 to 6 months and always after maintenance or service More frequent testing of the Labnet Excel electronic pipettor may be...

Page 15: ...should be filed with the transportation carrier In the event this product fails within the specified period of time because of a defect in material or workmanship contact Corning Customer Service at...

Page 16: ...isting of US medical devices regulatory classifications or specific information on claims visit www corning com resources www labnetlink com For a listing of trademarks visit www corning com clstradem...

Reviews: