Labnet Excel™ Single- and Multi-channel Electronic Pipettors
Instruction Manual
Catalog Numbers:
P3600L-10 P3608L-10 P3612L-10P3600L-20 P3608L-20 P3612L-20P3600L-200 P3608L-200 P3612L-200P3600L-1200 P3608L-1200 P3612L-1200
Page 1: ...l Single and Multi channel Electronic Pipettors Instruction Manual Catalog Numbers P3600L 10 P3608L 10 P3612L 10 P3600L 20 P3608L 20 P3612L 20 P3600L 200 P3608L 200 P3612L 200 P3600L 1200 P3608L 1200...
Page 2: ...Audible Signals 7 6 4 Blow out 7 6 5 Reverse Pipetting 7 6 6 Program Storing and Recall 7 7 0 AUTO Automatic Pipetting Mode 8 7 1 Setting Up the Volume in AUTO Mode Only 8 7 2 Setting Up the Volume in...
Page 3: ...lly explosive environment or with potentially explosive chemicals When using organic solvents or corrosive chemicals please check their compatibility with the pipet tips and the Labnet electronic pipe...
Page 4: ...5 to 0 5 P3612L 200 LE200 12 10 200 1 3 0 to 0 6 1 0 to 0 15 P3612L 1200 LE1200 12 100 1200 1 3 0 to 0 5 0 8 to 0 2 NOTE Labnet Excel electronic pipettor accuracy and precision specifications have bee...
Page 5: ...pipettor will automatically perform a calibration routine Figure 2 Insert the battery connector into the bottom right recess first then place the battery into position Figure 1 Open the battery cover...
Page 6: ...accuracy and precision specifications for the Labnet Excel electronic pipettors are guaranteed only when using Labnet BioFree tips Contact Corning Customer Service for pipet tip ordering information T...
Page 7: ...mode selection and volume settings This feature can speed up the setting process Enter Button Confirms your choice Count No Prog No Volume L Down Button Decreases volume and changes parameters Up But...
Page 8: ...ion map 6 0 SYS System Mode SYS mode is used to set up the function settings before pipetting Once the user has selected and stored these settings they will no longer need to set up these settings bef...
Page 9: ...tomatic reverse pipetting is to reserve the final drop of sample The Labnet Excel electronic pipettor will not perform reverse pipetting in MIX or SA modes 1 The Reverse Pipetting symbol should now be...
Page 10: ...s the Plunger button to aspirate the sample Press the Plunger button again to dispense the sample and the LCD will display the mixing cycle count and volume Press the Plunger to mix and dispense the s...
Page 11: ...lashes 3 Press the Enter button to confirm the selection 4 Use the Up Down button until the SA symbol flashes 5 Press the Enter button to confirm the selection 6 Under the COUNT symbol 01 will appear...
Page 12: ...y for dispensing 11 The pipettor will make two beeping sounds to remind the user of the last measured volume to dispense At the end of the sequential dispensing cycle if Reverse Pipettor mode is selec...
Page 13: ...on assembly downward until the cylinder lock appears Cylinder lock C Push the metal cylinder lock downward until you hear a click sound D The lower piston assembly will automatically come off easily F...
Page 14: ...the performance of your Labnet Excel electronic pipettor regularly e g every 3 to 6 months and always after maintenance or service More frequent testing of the Labnet Excel electronic pipettor may be...
Page 15: ...should be filed with the transportation carrier In the event this product fails within the specified period of time because of a defect in material or workmanship contact Corning Customer Service at...
Page 16: ...isting of US medical devices regulatory classifications or specific information on claims visit www corning com resources www labnetlink com For a listing of trademarks visit www corning com clstradem...