6 – LABNET BIOPETTE PLUS PIPETTORS
Nominal
Volume
(μL)
Volume
(μL)
Accuracy
(%)
Precision
(%)
Non-filter
Tips (μL)
8-channel and 12-channel Pipettors
8 - 10
12 - 10
Min.
0.5
±10.0
≤8.0
10
1
±8.0
≤6.0
5
±4.0
≤2.0
Max.
10
±2.0
≤1.2
8 - 50
12 - 50
Min.
5
±4.0
≤2.5
200
25
±3.0
≤1.2
Max.
50
±1.6
≤0.6
8 - 200
12 - 200
Min.
20
±3.0
≤1.5
200
100
±1.5
≤0.8
Max. 200
±1.0
≤0.6
8 - 300
12 - 300
Min.
50
±1.6
≤1.5
300
150
±1.2
≤1.0
Max. 300
±1.0
≤0.6
The accuracy and precision were obtained gravimetrically
using manufacturer tips performing at least 10 measurements
of distilled water at a temperature of 20°C ± 1°C according to
EN ISO 8655 standards.
The use of tips from other manufacturers or filter tips may
result in incorrect liquid aspiration and require pipettor
recalibration.
The design of the pipettor enables the user to recalibrate it
according to the information presented in Section 9.