Labnet BioPette PLUS BPP2 Instuction Manual Download Page 10

If you find an increase in the pipetting force, which
may happen after repetitive autoclaving of the pipette:

Remove the tip ejector. Unscrew the shaft nut, and
then remove the shaft, piston assembly, seal and 
O-ring. Rinse the removed parts in distilled water and
dry. Lubricate the internal surfaces of the seal and 
O-ring with lubricant that has been included with each
pipette. Reassemble the pipette in opposite order.

Note: All parts of the pipette can be autoclaved (see
Section 10)
The shaft of the 5000 and 10000 models should be
autoclaved without the filter. 

If the problem continues after carrying out the above
steps, contact your 

Labnet

representative.

Before returning the pipette, please ensure that the
pipette is completely free of any chemical, radioactive or
microbiological contamination which could pose a threat
during transport and repair. As far as it is possible, clean
the pipette.

10 - CLEANING AND STERILIZATION

Cleaning

External surfaces of the pipetting pushbutton, the ejector
pushbutton, the handgrip, the shaft nut and the adjust-
ment knob may be cleaned using a cloth dampened in
isopropyl alcohol. The remaining parts removed from the
pipette during pipette disassembly may be washed with
distilled water or isopropyl alcohol.

Sterilization:

The pipette can be sterilized in an autoclave at 121

°

C for

20 minutes. After sterilization, the pipette should be dried
and cooled to room temperature.

It is recommended:

- to sterilize the pipettes in autoclave with an initial vacu-

um and drying cycle,

- prior to sterilization unscrew the shaft nut slightly in the

BPP2-BPP1000 pipettes, and unscrew the shaft slightly
in the BPP5000 and BPP10000. After autoclaving these
parts should be screwed tight again.

- to set the locking ring in lower (unlocked) position prior

to sterilization.

The precision of the results should not alter if the pipet-
ting process and autoclaving are carried out as
described in this manual. Because a slight change in the
accuracy of the dosage may occur, it is recommended to: 
- check the calibration of the pipette after the initial first,

third and fifth autoclaving cycles and then after every 10
autoclaving cycles. 

11 - PIPETTE KIT AND ACCESSORIES

Pipette kit:

The pipettes are delivered in the kits including: 
• Pipette
• Instruction manual 
• Calibration key
• Shelf Clip
• Ejector regulation spacers 

(for pipette models BPP2 - BPP1000)

• Ejector cap (for pipette models BPP5000, BPP10000)
• Identification labels
• Filters (for pipette models BPP5000, BPP10000)
• Lubricant
The shelf clip assembly diagram is shown in Fig. 7. 

Accessories:

13

14

Description

10 μl tips, racked

200 μl tips, racked

1000 μl tips, racked

5000 μl tips, bulk pack

10000 μl tips, bulk pack

Carousel stand for 6 pipettes

Linear stand for 6 pipettes

Cat. No.

P1000-10-R

P1000-200-R

P1000-1000-R

P3998

K943

P3985

P3989

Summary of Contents for BioPette PLUS BPP2

Page 1: ......

Page 2: ...A N F G J I C A1 A2 B P C D E D L A B C D E F...

Page 3: ...P 5A 5B P 5 4 6 4D 4A 4B 4C 6A 2 1000 l 7 6B 5000 l 10000 l 6C 5000 l 10000 l D D D D D D...

Page 4: ...US is equipped with a built in tip ejector Fig 1D The construction of the ejector enables the user to adjust the length This allows for proper fit and ejection of nearly every tip type available When...

Page 5: ...PP2 Red figures at the bottom 1 100 l Increment 0 002 l Pipettes BPP10 BPP20 Red figures at the bottom 1 10 l Increment 0 02 l Pipettes BBPP100 BPP200 Black figures only 1 l Increment 0 2 l Pipettes B...

Page 6: ...touching the orifice of the tip Dispensing Place the end of the tip against the inside wall of the vessel at an angle of 10 to 40 degrees Press the pushbutton smoothly to the first stop Fig 2C Wait on...

Page 7: ...s are used for volumes between 0 1 l and 10 l They are used with the BPP2 and the BPP10 models Tips 200 These tips are used for volumes between 2 l and 200 l They are used with the BPP20 BPP100 and th...

Page 8: ...e 20 25 C and stabilized during weighing within 0 5 C Measurements should be conducted using distilled water Balance sensitivity should be suitable for the volume to be controlled 10 Recalibration pro...

Page 9: ...f the shaft is cracked or scored Remove the tip ejector Unscrew the shaft nut inspect the shaft and the piston assembly Replace the damaged parts see Section 12 When reasembling the pipette the nut sh...

Page 10: ...n an autoclave at 121 C for 20 minutes After sterilization the pipette should be dried and cooled to room temperature It is recommended to sterilize the pipettes in autoclave with an initial vacu um a...

Page 11: ...efore returning a pipette to a Labnet representative please ensure that the pipette is completely free of any contamination chemical microbial or radioactive Contact Labnet at 1 732 417 0700 fax 1 732...

Reviews: