Instruction Manual – 3

 A.  Pipetting  pushbutton: 

Consists of 2 parts: 

the pushbutton (A1) and the knob (A2).

 B.  Volume adjustment knob: 

Used to set the volume.

 C.  Shaft: 

Manufactured using high quality plastic, 

ensuring high chemical and mechanical strength.

 D.  Tip  ejector: 

For multi-channel pipettors, the tips are 

ejected sequentially, reducing the required force.

 F.  Shaft nut
 N.  Ejector  pushbutton
 R.  Multi-channel manifold: 

Contains a set of plungers 

and a set of flexible-suspension shafts, which function 

to reduce the force required to attach the tip.

Model Identification

The volume range of the pipettor is shown on the 

pipetting pushbutton.

 4.  Safety  Recommendations 

Long-term use of the pipettor relies on correct method of use. 

Please read and follow the instructions for use carefully. 
Symbols used: 

Danger, risk of injury.

NOTE

Risk of damage to the pipettor or errors in pipetting.

NOTE:

•  The pipettor is designed for the transfer of liquids only 

using the tip. Do not aspirate liquids without the tip 

attached. The aspirated liquid should not enter the 

pipettor, as it may cause damage.

•  Single-use tips reduce the risk of contamination of 

samples.

•  Keep the pipettor clean, avoiding the use of abrasive or 

corrosive cleaning agents (e.g., acetone). 

•  Keep the pipettor upright when there is liquid in the tip.
•  Only using the pipettor in accordance with the 

manufacturer’s instructions ensures the correct pipettor 

parameters are maintained.

Summary of Contents for BioPette A Series

Page 1: ...Labnet BioPette A Single channel 8 channel and 12 channel Pipettors Instruction Manual This manual is available in additional languages at www labnetlink com...

Page 2: ...mendations 3 5 Specifications 4 6 Pipettor Operation 6 7 Operating Instructions 10 8 Aspiration and Dispensing Instructions 11 9 Checking Pipetting Accuracy Parameters and Pipettor Recalibration 13 10...

Page 3: ...0 L and 50 300 L Nominal Volume L Cat No Range of Pipettor Volumes L Single channel Pipettors 2 P3960 2A 0 1 2 10 P3960 10A 0 5 10 20 P3960 20A 2 20 100 P3960 100A 10 100 200 P3960 200A 20 200 1 000 P...

Page 4: ...5000A and P3960 10000A models 1 Filters for P3960 5000A and P3960 10000A models 3 Filters The 5 000 and 10 000 L pipettors include a replaceable filter fitted at the bottom of the shaft which prevents...

Page 5: ...pipetting pushbutton 4 Safety Recommendations Long term use of the pipettor relies on correct method of use Please read and follow the instructions for use carefully Symbols used Danger risk of injury...

Page 6: ...work in the laboratory Take special care when pipetting aggressive substances Use appropriate protective attire e g clothing goggles and gloves Avoid pointing the pipettor at yourself or others durin...

Page 7: ...0 20 200 Min 20 1 2 0 60 200 100 0 8 0 25 Max 200 0 6 0 20 1 000 Min 100 1 6 0 40 1 000 500 0 7 0 20 Max 1 000 0 6 0 15 5 000 Min 1 000 0 6 0 25 5 000 2 500 0 6 0 20 Max 5 000 0 5 0 15 10 000 Min 1 0...

Page 8: ...el number displayed below counter window Section 5 Position the pipettor vertically when attaching tips Single channel pipettors Push the pipet tip on firmly using a slight twisting motion to ensure a...

Page 9: ...ress the tip ejector pushbutton Slide the ejector from the arbor 2 Tip ejector assembly Press the tip ejector pushbutton Line the metal tip ejector up with the arbor Slide the ejector on the arbor 3 T...

Page 10: ...tal arbor clockwise to secure the ejector and adjust its length NOTE Leave a space of at least 1 mm between the shaft and tip ejector collar 3 Tip ejector adjustment Remove the ejector pushbutton Use...

Page 11: ...inted on the bottom counter drum Examples of counter indications The decimal point in the volume of aspirated liquid L or mL is indicated by change in color of digits Black digits represent integer nu...

Page 12: ...The pipet tip should be changed if a droplet remains on the end of the tip from the previous pipetting operation Each new pipet tip should be pre rinsed with the liquid to be pipetted Liquid should ne...

Page 13: ...one second and then withdraw the pipet tip from the liquid CAUTION Do not touch the used tip Dispensing Liquid 3 Place the end of the pipet tip against the inside wall of the vessel at an angle of 10...

Page 14: ...n a new pipet tip is used NOTE Normally the degree of error resulting from viscous liquids is negligible if pipetting is performed slowly and carefully however this can be minimized further by holding...

Page 15: ...ad sterilization or autoclave processes and frequency of replacement of component parts If during pipettor operation the accuracy error the difference between the real aspirated volume and the preset...

Page 16: ...y mg 0 1 V 10 0 001 10 V 100 0 01 100 V 1 000 0 1 V 1 000 0 1 When calculating the liquid volume aspirated by the pipettor the conversion factor Z L mg for distilled water or a liquid with comparable...

Page 17: ...06 0 63 0 0262 100 10 9 84 10 16 2 50 0 104 200 20 19 76 20 24 6 30 0 262 1 000 100 98 4 101 6 25 00 1 04 5 000 500 494 506 125 00 5 2 10 000 1 000 975 1 025 250 00 10 4 8 channel and 12 channel Pipe...

Page 18: ...se the aspirated volume Use the values given in the above table to precisely adjust the volume 4 Remove the key and replace the pipetting pushbutton by first replacing the knob then the pushbutton Det...

Page 19: ...which form components of the pipettor PP PC POM PA PPS PVDF Sterilization Sterilization using an autoclave The pipettor can be sterilized in an autoclave at 121 C for 20 minutes Sterilization under o...

Page 20: ...pipet tip is emptied too fast Decrease the speed of pressing the pipettor pushbutton The pipet tip wettability has increased due to extensive use Replace the tip with a new one Droplets of air appear...

Page 21: ...w the shaft set wash the parts Replace the elements with new ones if necessary Apply a small amount of lubricant on the plunger and reassemble the set in the correct order The inside of the pipettor i...

Page 22: ...arts are depicted below NOTE The replacement of the plunger requires conducting the calibration procedure according to Section 9 A1 N F G J I C D L A2 Item Description Model L Cat No Qty Pk A Pipettin...

Page 23: ...0 SP19443 1 1 000 SP29419 1 5 000 SP19452 1 10 000 SP19453 1 F Shaft nut 2 1 000 SP29313 1 G Piston assembly with spring and bushing 2 SP19381 1 10 SP19382 1 20 SP19383 1 100 SP19384 1 200 SP19385 1 1...

Page 24: ...if the product is used for its intended purpose and within the guidelines specified in the supplied instructions manual This warranty does not cover damage caused by accident neglect misuse improper...

Page 25: ...ke a note of the model number serial number date of purchase and supplier here Model No _______________________________________________ Serial No ________________________________________________ Date...

Page 26: ...PAC I FIC Australia New Zealand t 61 427286832 Chinese Mainland t 86 21 3338 4338 f 86 21 3338 4300 India t 91 124 4604000 f 91 124 4604099 Japan t 81 3 3586 1996 f 81 3 3586 1291 Korea t 82 2 796 95...

Reviews: