3.7 Power Save Mode
If the unit is left On but inactive for long periods, it will enter a power saving mode. These modes are described below.
Display
Time Frame
Temperature
STND_BY
1 hour of inactivity
60°C
PWRSAVE
3 hours of inactivity after entering STND_BY mode
Ambient (heater Off)
4.0 Troubleshooting
Error
Possible Cause
Solution
Plates not sealing correctly.
Insufficient sealing time.
Check with the sealing film manufacturer for
recommended sealing time and temperature.
Sealing temperature is too low.
Top of the plate is not flat.
Turn the plate 180°, and reseal.
Sealing film is curling and
moving out of position.
Temperature difference and electrostatic
charge affecting the seal.
Use the sealing frame to secure the seal in
place.
Sealing film is sticking to
platen.
Foreign material stuck to the platen.
Follow the procedure for cleaning the heating
platen.
4.1 Error Codes
Code
Possible Cause
Solution
Error 1
Tray motor or switch failure.
Reboot the unit.
Error 2
Error 3
Heater motor or switch failure.
Error 4
Error 5
Heater is not working or temperature sensor (PT-1000) failure.
Error 6
Heater PCBA failure.
Error 8
Heater is not working or temperature sensor (PT-1000) failure.
Error 9
Safety door/housing switch was activated during drawer operation.
Foreign object is stuck between the tray door and the tray plate.
Turn Off the unit. Remove the foreign
object, and reboot the unit.