  Instruction Manual        | 

         3

6.1 Heating and Countdown Timer

1.  Turn the unit On using the Main Power switch.
2.  Press the Up or Down buttons to set the block to the desired temperature. Holding the Up or Down button for more the eight 

seconds will increase or decrease the set point rapidly. The display will automatically return to current temperature four seconds 
after the switch is released. 

NOTE:

 The Timer may not be adjusted with the unit running.

3.  Press the Mode button until 1:00 is displayed. This is the default Count Down timer setting. The maximum time is setting is 19:59.
4.  Press the Up or Down button to set the desired countdown time.
5.  Press START/STOP to begin heating.
6.  The LCD will alternate between actual temperature and time remaining.
7.  When time runs out, the LCD will alternate between the current temperature and “---“. The beeper will sound when “---“ is displayed.
8.  Heating will NOT turn Off when time expires.
9.  To cancel the countdown timer, press the Mode button one time to advance to the standard temperature only read out.

 7.0  Calibration

1.  Press and hold START/STOP then simultaneously power up the unit with the ON/OFF switch.
2.  Release START/STOP when the LCD displays “---“.
3.  The beeper will sound 4 times, and the display will show the calibration set point, and the “:” will flash rapidly.
4.  Press the Up or Down button to set the desired temperature at which you want to calibrate the unit.
5.  Press START/STOP to begin calibration. Heat LED will blink On and Off when the unit is in active temperature control.
6.  Allow 10 to 40 minutes for the unit to heat up and reach equilibrium. A high temperature calibration set point will require more time.
7.  After the unit has equalized, the LCD will flash “---“ and the current temperature.
8.  Read the thermometer (or sensor meter) and use the Up and Down buttons to adjust the display to the thermometer or sensor reading.
9.  Press START/STOP to save the new settings.
10. Press START/STOP again. The unit will adjust to the operating temperature with the new calibration factor. Allow sufficient time 

for the unit to re-equilibrate, and then compare the thermometer or sensor reading to the calibrated display. They should closely 
match (± 0.2°C). If not, repeat calibration procedure.

 8.0  Troubleshooting  Guide

Problem

Solution

Display/LEDs do not light up

 



Check power cord and outlet.

 



Check ON/OFF switch.

 



Call for service.

Unit not heating

 



Check if the set point is more than 5°C above room temperature.

 



Check the heating LED.

 



Press START/STOP.

 



Call for service.

Block or Sample temperature is not 
the same as the display temperature   

 



Check if the unit is in heating mode.

 



Check if the unit is sitting in a draft.

 



Check accuracy of thermometer.

 



Check if the reference thermometer is making good contact.

 



Follow the Calibration procedure.

 9.0  Cleaning and Maintenance

Make sure that the dry bath and block are cool and the power cord is disconnected before performing any cleaning or maintenance. 
Repair or maintenance should only be performed by an authorized service technician. 

The dry bath may be cleaned with a moist cloth containing a mild soap solution. Do not immerse the dry bath in water or any liquid. 

Spills: In the event liquid is accidentally spilled into the block area, disconnect the plug from the outlet and turn the unit upside down 
to minimize liquid contact with the internal components. Allow the unit to dry. Have a qualified service technician clean the unit and 
replace any damaged parts. 