Labnet 311D Instruction Manual Download Page 5

  Instruction Manual        | 

         3

5.4 Leveling

The unit must sit level and solidly. Leveling feet are supplied and must be installed in the four holes in the bottom corners of the unit. 
With the feet installed and the unit standing upright, each foot can be raised by turning it in a counterclockwise direction. Adjust the 
foot at each corner until the unit stands level and solid without rocking. If the unit must be moved, turn the leveling feet all the way 
clockwise to prevent damage while moving.

5.5 Cleaning

The unit chamber should be cleaned and disinfected before use. Remove all of the interior parts, and clean thoroughly with a 
disinfectant that is appropriate to your application. 

NOTE:

 Do not use chlorine-based bleaches or abrasives as this will damage the 

chamber interior. Regular periodic cleaning is required.

 6.0  Controls

1.  Power switch: 

The main power ON/OFF switch controls all power to the unit. The power must be On before any systems are 

operational. The green switch will be lighted when ON.

2.  Main temperature control: 

This control consists of the digital display and Up/Down buttons for inputting set point temperatures 

and calibration.

3.  Safety thermostat: 

The Safety is completely independent of the Main Controller. The Safety guards against any failure of the Main 

Controller that would allow temperature to rise past set point. If temperature rises to the safety set point, the Safety takes control 
of the heating element and allows continued use of the incubator until the problem can be resolved or service can be arranged.

4.  Fan switch: 

This operates the turbo fan inside the chamber for fast heat up times and recovery. The switch is lighted when 

activated.

5.  Safety light: 

This pilot light comes on when the Safety thermostat is activated. Under normal operating conditions this pilot light 

should never be on.

 7.0  Operation

Check power supply against unit serial plate. They must match. Plug the power cord into the electrical outlet. 

Push power switch ON, and turn the Safety thermostat to its maximum position, clockwise.

7.1 Set Main Temperature Control

Enter desired set point temperature. To enter set point mode on the control, press either the Up or Down button one time. The digital 
display will start to blink, going from bright to dim. While blinking, the digital display is showing the set point. To change the set 
point, use the Up and Down buttons. If the buttons are not pressed for five (5) seconds, the display will stop blinking and will read 
the temperature of the unit. 

NOTE:

 The Safety thermostat should be turned to its maximum position, until the unit has stabilized at 

desired set point temperature. Allow the incubator at least 24 hours to stabilize.

7.2 Calibration

The Main Temperature Controller was calibrated at the factory at 37°C; however, it is recommended that calibration is done 
once the unit is installed in its working environment and has been stable at set point for several hours. Place a certified reference 
thermometer in the chamber. Be certain the thermometer is not touching any shelving. Allow the temperature to stabilize again 
until the thermometer reads a constant value for one hour. Compare the digital display with the reference thermometer. If there is 
an unacceptable difference, put the display into calibration mode by pressing both the Up and Down arrow buttons at the same time 
until the two outside decimal points begin to flash. While the decimal points are flashing the display can be calibrated by pressing 
the Up or Down arrow buttons until the display reads the correct value. Allow the incubator temperature to stabilize again, and 
recalibrate if necessary.

7.3 Set Safety Thermostat

As mentioned above, the Safety thermostat should be initially set to its maximum position, to allow the unit to stabilize. Once the 
incubator is stabile at the desired set point, turn the Safety thermostat counterclockwise until the Safety light turns on. Next, turn 
the thermostat clockwise just until the Safety light turns off. Then turn the thermostat clockwise two (2) of its smallest divisions on 
its scale past the point where the Safety light went out. This will set the Safety thermostat at approximately 1°C above the main 
temperature set point.

Summary of Contents for 311D

Page 1: ...Labnet 311D and 611D Digital Incubators Instruction Manual Catalog Numbers I 5311 D I 5611 D...

Page 2: ...2 5 3 Lifting Handling 2 5 4 Leveling 3 5 5 Cleaning 3 6 0 Controls 3 7 0 Operation 3 7 1 Set Main Temperature Control 3 7 2 Calibration 3 7 3 Set Safety Thermostat 3 7 4 Fan Switch 4 8 0 Cleaning an...

Page 3: ...g of materials is done at approximately atmospheric pressure and no flammable volatile or combustible materials are being heated These units are not intended for hazardous or household locations or us...

Page 4: ...equipment If you have any questions about local requirements please contact the appropriate local agency Installation can be performed by the end user 5 1 Power Source Check the data plate for voltage...

Page 5: ...at to its maximum position clockwise 7 1 Set Main Temperature Control Enter desired set point temperature To enter set point mode on the control press either the Up or Down button one time The digital...

Page 6: ...too low see Section 7 3 Controller set too low see Section 7 1 Unit not recovered from door opening wait for display to stop changing Unit not recovered from power failure or being turned off incubat...

Page 7: ...9 3 Other Code Description Solution Other 1 Controller on at all times locked up Turn unit off and on to reset If cannot change any condition on the front panel call Customer Service Other 2 Front pan...