OilSET-1000 Oil Separator Alarm Device 

 

     

                                 

D15492AE-2 

Installation and Operating Instructions

  

 

 

Copyright © 2019 Labkotec Oy 

9/13 

We reserve the right for changes without notice 

REPAIR AND SERVICE 

The sensor should be cleaned and the operation should also be tested 
when  emptying  the  oil  storage  chamber  or  at  least  once  every  six 
months. The easiest way to check the operation is to lift the sensor up 
in the air and to put it back to the separator. The operation is described 
in chapter 3. 

For cleaning, a mild detergent (e.g. washing-up liquid) and a scrubbing 
brush can be used. 

The mains fuse (marked 125 mAT) can be changed to another glass 
tube fuse 5 x 20 mm / 125 mAT complying IEC/EN 60127-2/3. Any other 
repair  and  service  works  on  the  device  may  be  carried  out  only  by  a 
person who has received training in Ex-i devices and is authorized by 
the manufacturer. 

In case of queries, please contact Labkotec 

Oy’s service. 

SAFETY INSTRUCTIONS 

OilSET-1000 control unit must not  be  installed  in  potentially  explosive 
atmosphere. Sensors connected to it may be installed in zone 0, 1 or 2 
potentially explosive atmospheres. 

In  case  of  installations  in  explosive  atmospheres  the  national 
requirements  and  relevant  standards  as 

IEC/EN  60079-25  and/or 

IEC/EN 60079-14

 must be taken into account. 

Warning!  If  the  cabling  is  voltage-tested,  the  sensor  must  be 
disconnected. 

If  electrostatic  discharges  can  cause  hazards  in  the  operating 
environment,  the  device  must  be  connected  into  equipotential  ground 
according  to  requirements  with  regards  to  explosive  atmospheres. 
Equipotential grounding is made by connecting all conductive parts into 
same potential e.g. at the cable junction box. Equipotential ground must 
be earthed. 

The device does not include a mains switch. A  two pole mains switch 
(250 VAC 1 A), which isolates both lines (L1, N) must be installed in the 
main power supply lines in the vicinity of the unit. This switch facilitates 
maintenance and service operations and it has to be marked to identify 
the unit. 

When executing service, inspection and repair in explosive atmosphere, 
the rules in standards 

IEC/EN  60079-17 and IEC/EN  60079-19

 about 

instructions of Ex-devices must be obeyed. 

 

 

Summary of Contents for OilSET-1000

Page 1: ...8 29 006 260 Fax 358 29 006 1260 23 9 2019 Internet www labkotec fi D15492AE 2 1 13 Copyright 2019 Labkotec Oy We reserve the right for changes without notice OilSET 1000 Oil Separator Alarm Device wi...

Page 2: ...trol unit 4 2 2 SET DM3AL sensor 5 2 3 Cable joint 5 3 OPERATION AND SETTINGS 6 3 1 Operation 6 3 2 Factory settings 7 4 TROUBLE SHOOTING 8 5 REPAIR AND SERVICE 9 SAFETY INSTRUCTIONS 9 6 TECHNICAL DAT...

Page 3: ...is installed into the light liquid storage chamber and gives an alarm when the chamber is filled to a pre determined degree The sensor is normally immersed in water The function is based on the measu...

Page 4: ...re beneath the mounting holes of the front cover The connectors of the external conductors are isolated by separating plates The plates must not be removed The plate covering the connectors must be in...

Page 5: ...ined in figure 3 Cable shields and possible excess wires need to be connected to the same point in galvanic contact Please make sure that the sensor and cable between OilSET 1000 control unit and the...

Page 6: ...LED indicators are off Relays 1 and 2 are energized Oil alarm SET DM3AL sensor is immersed in oil The sensor gives an alarm when the upper electrode is in oil Mains LED indicator is on Oil Alarm LED...

Page 7: ...OilSET 1000 is not as described in the previous chapter check that the device settings are as in figure 7 Change the settings according to the following instructions if needed The following tasks mus...

Page 8: ...ensor should give an alarm again Problem FAULT LED indicator is on Possible reason Current in sensor circuit too low cable break or too high cable in short circuit The sensor might also be broken To d...

Page 9: ...tentially explosive atmosphere Sensors connected to it may be installed in zone 0 1 or 2 potentially explosive atmospheres In case of installations in explosive atmospheres the national requirements a...

Page 10: ...e g 19 2019 Due to non linear characteristics of the sensor voltage the interaction of both capacitance and inductance must be taken into account The table below indicates the connecting values in exp...

Page 11: ...DECLARATION OF CONFORMITY...

Page 12: ...DECLARATION OF CONFORMITY...

Page 13: ...DECLARATION OF CONFORMITY...

Reviews: