Labconco WaterPro RO 90750 Series User Manual Download Page 31

Chapter 5: Maintaining Your WaterPro RO System 

 

27 

Original Instructions 

6.

 

Fill the carbon/prefilter bowl approximately ¾ full with the cleaning solution. 

7.

 

Reinstall the bowl. 

8.

 

Open the feed water valve. 

9.

 

Plug the unit into an electrical outlet, activate the unit and allow it to operate for 5 minutes. 

10.

 

Disconnect the unit from the electrical outlet and allow the membrane to soak for at least 2 
hours. 

11.

 

Relieve the pressure in the system by connecting unit to an electrical outlet and pressing 
the dispense button. 

 

CAUTION:  Do not attempt to remove the acid-filled bowls when the system is 
pressurized

ATTENTION : N’essayez pas d’enlever les bols remplis d’acide lorsque le 
système est sous pression.

 

12.

 

Carefully remove the prefilter bowl and dispose of the water appropriately. 

 

CAUTION:  The water in the bowl contains cleaning solution. Handle and 
dispose of properly

ATTENTION : L’eau dans le bol contient de la solution de nettoyage. 
Manipulez-le et éliminez-le d’une manière dûment autorisée. 

 

13.

 

Install a new prefilter (first bowl from right) and carbon/prefilter (second bowl) cartridges. 

14.

 

Connect a hose to the dispensing valve and route to drain. Activate the unit by connecting 
the unit to an electrical outlet. Set time dispense for 60 minutes and direct water to drain. 

15.

 

At the end of the hour, set off feed-water valve and release the pressure by pressing the 
dispense button. 

16.

 

Start the unit. The RO module is now ready for normal operation. 

Organic Fouling 

Organic fouling of the membrane can be caused by silt, bacterial slime, or any other organic 
materials. 
 

 

 

The following recommended cleaners may be used: 

 

 

Sodium hydroxide 0.5% 

 

Sodium dodecyl sulfate 0.1% (Adjust pH to 11.5 with HC1, if necessary) 

 

Trisodium phosphate, 1% 

 

Sodium tri-polyphosphate, 1% 

 

Tetra sodium EDTA, 1% (Adjust pH to 11.5 with HC1, if necessary) 

For cleaning procedure, follow steps 1 through 17 under Inorganic Fouling. 

 

 

Summary of Contents for WaterPro RO 90750 Series

Page 1: ...WaterPro RO Systems Models 90750 Series...

Page 2: ...ween the hours of 7 30 a m and 5 30 p m Central Standard Time Part 9087305 Rev A ECO K940 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts...

Page 3: ...s 10 Component Identification 11 Setting Up the WaterPro RO System 13 Feed Water Connection 13 Tubing Installation 13 RO Membrane Installation 14 Optional Dispense Gun Installation 16 WaterPro PS and...

Page 4: ...TEM DIMENSIONS 37 APPENDIX C WATERPRO RO SYSTEM SPECIFICATIONS 38 RO Specifications 38 RO Production Rates Table 40 RO Performance Efficiency Table 41 APPENDIX D WATERPRO RO SYSTEM ACCESSORIES 42 Expe...

Page 5: ...Station to directly feed the Polishing Station with RO purified water It may also be connected to a Labconco Glassware Washer to dispense RO purified water for the final rinse cycles of the water An o...

Page 6: ...ing water quality in microsiemens S water temperature in C timed dispense duration from 0 99 9 minutes and a storage tank full indication light The RO System uses built in activated carbon and particu...

Page 7: ...ou des blessures s rieuses Throughout this manual potentially hazardous conditions are identified using the following words and symbols This is the safety alert symbol It is used to alert you to poten...

Page 8: ...Chapter 1 Introduction 4 Original Instructions...

Page 9: ...STEM SPECIFICATIONS for electrical and environmental conditions specifications and requirements Location WARNING Do not install the WaterPro RO System directly over or near equipment that uses electri...

Page 10: ...d tachable qui n a pas la tension nominale pour l unit The 115V WaterPro RO Station comes with a detachable 15 amp power cord which has a plug to connect to a standard outlet For model 9075020 the el...

Page 11: ...ge life span are directly related to the feed water It is important to establish feed water pressure and flow rate before installing and operating the unit If your feed water or feed water quality has...

Page 12: ...Chapter 2 Prerequisites 8 Original Instructions...

Page 13: ...r WaterPro RO System installation site is properly prepared you are ready to unpack inspect install and test your WaterPro RO System Read this chapter to learn how to Unpack and move your WaterPro RO...

Page 14: ...pping damage Do not discard the carton or packing material for your WaterPro RO System until you have checked all of the components and installed and tested the WaterPro RO System WaterPro RO System C...

Page 15: ...ge 10 Flush Valve The flush valve opens to flush the membrane and storage tank after 12 hours of inactivity 11 Dispensing Valve Manual dispensing valve dispenses RO water when the valve is open and au...

Page 16: ...Chapter 3 Getting Started 12 Original Instructions Component Identification 11 14 21 18 8 19 16 5 2 1 4 3 6 9 7 10 12 20 13 15 17 TO DRAIN Figure 2...

Page 17: ...he supply line a valve should be installed so the WaterPro RO may be isolated from the feed water supply when required Line pressure should not exceed 100 psi 7 Bar Feed water line must be able to sup...

Page 18: ...from the unit 5 Center the lower spacer disk over the support post in the bottom of the bowl NOTE Orient the disk in the bowl so that the slots are upward See Figure 4 6 Install the lower adapter int...

Page 19: ...Chapter 3 Getting Started 15 Original Instructions Figure 4...

Page 20: ...n the right side of the storage tank housing using the hardware provided See Gun Installation Instruction Sheet 6 Remove the dispense gun with the tubing from the shipping carton 7 Inside the cabinet...

Page 21: ...tubing required for connection to a Labconco WaterPro PS System or a Labconco Glassware Washer NOTE If no optional storage tank is connected to the RO system the Glassware Washer will require the unit...

Page 22: ...create a fire or electrical hazard Do not remove or service any electrical components without first disconnecting the RO system from the outlet Preserve RO membrane whenever unit is unplugged for more...

Page 23: ...s detailed in Chapter 3 Getting Started you are ready to begin using your WaterPro RO System Read this chapter to learn how to Understand the display Operate the unit If the unit is not operated as sp...

Page 24: ...peration and displays the water purity measured in conductivity a Microsiemens S This is the normal operating mode and displays the water purity measured in conductivity b Temp Displays temperature of...

Page 25: ...ve to drain Dispensing valve will close automatically the 17 liter storage tank will fill and unit will automatically shut off Unit is now ready to dispense RO water any time from the dispensing valve...

Page 26: ...electrical outlet Disconnect unit to service Preserve membrane if unit is left unplugged for more than a week Check to ensure security of drain lines Keep a log of RO performance Shut Down When chang...

Page 27: ...ystem requires little maintenance Read this chapter to learn The Maintenance Schedule Sanitizing the System Service Tips Maintenance Schedule As needed Change prefilter and carbon prefilter if more th...

Page 28: ...he system or preserving the membrane Avoid splashing the solution on skin or clothing Ensure that all piping connections are tight to avoid leakage Always depressurize the system COMPLETELY before dis...

Page 29: ...nect the unit by unplugging it from the outlet 4 Remove both 20 prefilter and 20 carbon prefilter cartridges Do not remove RO membrane 5 Carefully wash inside the prefilter bowl carbon prefilter bowl...

Page 30: ...rises noticeably Inorganic Fouling Inorganic fouling of the membrane is caused by mineral scale such as Calcium carbonate Calcium sulfate Barium sulfate Strontium sulfate Iron or manganese Inorganic...

Page 31: ...e and dispose of properly ATTENTION L eau dans le bol contient de la solution de nettoyage Manipulez le et liminez le d une mani re d ment autoris e 13 Install a new prefilter first bowl from right an...

Page 32: ...rmaldehyde in the RO water collected in Step 1 6 Install a new 20 prefilter cartridge catalog number 9021000 in the first bowl from the right 7 Fill the carbon prefilter bowl with the 0 1 formaldehyde...

Page 33: ...hut off feed water supply and release pressure in the system by pressing the dispense valve button 2 With the Spanner wrench provided remove the first bowl on the right to replace the Prefilter and th...

Page 34: ...Chapter 5 Maintaining Your WaterPro RO System 30 Original Instructions...

Page 35: ...ay Unit not plugged into outlet Plug WaterPro RO into appropriate electrical service Circuit breaker tripped fuse blown Reset circuit breaker replace fuse Power Switch is in the OFF position Press pow...

Page 36: ...25 C Install water temperature mixing valve Kinked or restricted tubing in system Trace restriction and remove or straighten tube Decrease in RO back pressure Increase back pressure relief valve to 1...

Page 37: ...ormal With increased inlet TDS RO S increases RO membrane scaling or other deposits Clean membrane per instructions in RO cleaning section Excessively high S RO membrane fouled Biofilm formed Clean RO...

Page 38: ...Chapter 6 Troubleshooting 34 Original Instructions...

Page 39: ...29100 Low Pressure Switch 4 1 1303200 High Pressure Switch 5 1 9047503 Pump Assembly115V 5a 1 9047502 Pump Assembly 230V 6 1 9106500 Valve Dump Dispense 115V Inlet 6a 1 9108900 Valve Dump Dispense 230...

Page 40: ...Appendix A WaterPro RO System Components 36 Original Instructions 1 9 8 7 6 5 4 3 2 10 12 11 13 6 17...

Page 41: ...A AP PP PE EN ND DI IX X B B W WA AT TE ER RP PR RO O R RO O S SY YS ST TE EM M D DI IM ME EN NS SI IO ON NS S 6 0 15 2 cm 8 6 21 8 cm 30 8 78 2 cm 7 6 19 3 cm 15 9 40 4 cm 35 3 89 7 cm 10 6 26 9 cm 3...

Page 42: ...S and or Labconco Glassware Washer connection From RO Up to 1 L min at 25 C 8 7 L min typical flow gravity from storage tank Optional Storage Tanks 70 liter with or without pump and dispensing gun 910...

Page 43: ...lowing conditions Indoor use Altitude up to 6562 Ft 2000 m Maximum relative humidity 80 for temperatures up to 88 F 31 C decreasing linearly to 50 relative humidity at 104 F 40 C Main supply voltage f...

Page 44: ...Appendix C WaterPro RO System Specifications 40 Original Instructions RO Production Rates Table...

Page 45: ...Appendix C WaterPro RO System Specifications 41 Original Instructions RO Performance Efficiency Table...

Page 46: ...Osmosis membrane and adapter package Required for operation 9062200 Replacement Carbon Prefilter contains 2 each 9067201 Prefilter Carbon Filter Kit required for operation Includes prefilters 3 carbon...

Page 47: ...lation and helps to protect the wall from moisture 9077400 Support Stand for converting wall mounted WaterPro RO System to bench mounted 9113200 WaterPro RO PS Mobile Stand allows the mounting of a RO...

Page 48: ...44 Original Instructions Operating Log...