22
5.1 Typical Wiring Diagram
with
Optional Temperature Sensor
Note :
Temp Sensor
Optional
Temp sensor connects to Input 3.
Input 3 has to be set to analogue and function set to Temperature.
REFER TO
INSTALLATION
INSTRUCTIONS
BEFORE
CONNECTING
www.tel-uk.com
Made in England
TEL
TO REMOTE
RELAY
INTERFACE BOX
I/P1
I/P3
I/P2
15V
PSU
I/P
-
+
0-10V
O/P
1
R1
R2
R3
INPUT 2
VOLT FREE RELAY
CONTACT INPUT
INPUT 1
VOLT FREE RELAY
CONTACT INPUT
TEMPERATURE
SENSOR
120VAC POWER
SUPPLY
(15VDC Output)
SM6 AIRFLOW
SENSOR
RJ11 PLUG IN
CABLE (2M / 6’)
R1,R2,R3
VOLT FREE RELAY
OUTPUTS
Max 5A 250VAC / 30VDC
0-10V RETRANSMISSION
OUTPUT
(OPTIONAL)
PLUG IN COMS
CABLE (2M / 6’)
RS485 COM PORT Connector
Modbus RTU or BACnet
(OPTIONAL)
GND
(A)+
(B)-
+VE
To PC or BMS
To OTHER AFA1000
UNITS
bk
gn
red
-
sw
+
sw
+
+