Chapter 3: Getting Started

 

 

Product Service 816-333-8811 or 1-800-522-7658

 

11

If you did not receive one or more of the components listed for your Freeze Dryer, 
or if any of the components are damaged, contact Labconco Corporation 
immediately for further instructions. 

Setting Up Your Freeze Dryer 

After you verify receipt of the proper components, move your Freeze Dryer to the 
location where you want to install it.  Then, follow the steps listed below. 
 

Component Orientation 

 
The refrigeration system in the Freeze Dryer draws air in through the rear of the 
cabinet and exhausts air out through the sides.  A minimum of 3" should be 
allowed between the back and sides of the Freeze Dryer and the adjacent wall 
surface. Restriction of the airflow through the cabinet during operation could 
adversely affect performance. 
 
After the console model is positioned at the location where it will be operated, it 
should be securely attached to the floor.  Use (2) 1/4" x 5.0 long wood screws or 
lag bolts if the floor is wood or concrete anchors if the floor is concrete.  Two .312 
diameter holes are located in the floor of the lower cabinet.  Using these holes as a 
template, mark the floor directly under the two holes.  Roll the Console Freeze 
Dryer aside and drill suitable pilot holes for the selected screws/anchors. Install the 
vacuum pump prior to securing the Console Freeze Dryer to the floor.  Roll the 
Console Freeze Dryer back into position and install the screws/anchors through the 
holes in the cabinet floor and into the building floor. 
 

FAILURE TO ATTACH THE CONSOLE FREEZE DRYER TO 
THE FLOOR COULD RESULT IN ACCIDENTAL TIPPING 
AND POSSIBLE INJURY 

 

Vacuum Pump Connection 

 
A vacuum pump as described in 

Chapter 2: Prerequisites

 is required to operate 

your Freeze Dry System properly.  The Freeze Dryer is equipped with a 3/4" ID, 
heavy wall, vacuum hose for connecting the collector chamber to the vacuum 
pump. 
 

Make certain that the oil in the vacuum pump is at the proper 
level. 

 
Place the vacuum pump near the benchtop Freeze Dryer models.  For console 
models, from the back of the unit, place the vacuum pump on the floor of the 
cabinet.   
 

 

!

 

)

 

Summary of Contents for FreeZone 76700 Series

Page 1: ...tes complete your product registration card online at register labconco com User s Manual Please read the User s Manual before operating the equipment Labconco Corporation 8811 Prospect Avenue Kansas...

Page 2: ...addressed in this manual or if you need technical assistance contact Labconco s Customer Service Department or Labconco s Product Service Department at 1 800 821 5525 or 1 816 333 8811 between the hou...

Page 3: ...GETTING STARTED 8 Unpacking Your Freeze Dryer 9 Freeze Dryer Components 9 Setting Up Your Freeze Dryer 11 Component Orientation 11 Vacuum Pump Connection 11 Electrical Connection 12 Drying Chamber or...

Page 4: ...TROUBLESHOOTING 32 Vacuum Pump 32 Gaskets Tubing Connections Sample Valves 33 System Components Collection Chamber Isolation 35 Refrigeration Module Operation 37 CHAPTER 8 MODIFYING YOUR FREEZE DRYER...

Page 5: ......

Page 6: ...ively used in the development of pharmaceuticals e g antibiotics and preservation of biologicals e g proteins plasma viruses and cell lines The nondestructive nature of this process has been demonstra...

Page 7: ...on the fact that the heat of sublimation is usually absorbed on one side of the frozen sample and must travel through the frozen layer to vaporize water at the other surface In addition as the sample...

Page 8: ...ystem With any given instrument the capacity is based on the surface area of the sample the eutectic temperature and concentration of the sample and the rate and amount of heat transferred to the froz...

Page 9: ...intenance and making minor modifications to your Freeze Dryer are also included Chapter 1 Introduction provides a brief overview of the Freeze Dryer explains the organization of the manual and defines...

Page 10: ...y the exclamation icon Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to your Freeze Dryer Important information is presented...

Page 11: ...let requires a 20 Amp circuit breaker or fuse for models rated at 115V 60 Hz An outlet equipped with a 15 Amp circuit breaker or fuse is required for models rated at 230V 50 60 Hz The power cord on 11...

Page 12: ...table for 3 4 ID vacuum hose which is provided It is recommended that the vacuum pump is equipped with an exhaust filter to minimize oil mist exhausting from the vacuum pump The operating vacuum level...

Page 13: ...connect the electrical supply source to your Freeze Dryer properly exhaust your Freeze Dryer safely use solvents with your Freeze Dryer The FreeZone 2 5 Liter Benchtop Freeze Dryer System weighs over...

Page 14: ...CEPTED IF YOUR FREEZE DRYER WAS DAMAGED IN TRANSIT YOU MUST FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER LABCONCO CORPORATION AND ITS DEALERS ARE NOT RESPONSIBLE FOR SHIPPING DAMAGE DO NOT DISCARD T...

Page 15: ...enchtop Cascade Freeze Dryer 230 60 X 7670041 2 5L Benchtop Cascade Freeze Dryer 230 60 X X 7670060 2 5L Benchtop Cascade Freeze Dryer 230 50 X 7670061 2 5L Benchtop Cascade Freeze Dryer 230 50 X X 76...

Page 16: ...wood screws or lag bolts if the floor is wood or concrete anchors if the floor is concrete Two 312 diameter holes are located in the floor of the lower cabinet Using these holes as a template mark th...

Page 17: ...cuum pump power cord if necessary Electrical Connection Plug the power cord into the receptacle on the back of the Freeze Dryer and plug the other end into a suitable power receptacle Drying Chamber o...

Page 18: ...damage to the vacuum pump Sugars and proteins typically will have minimal negative effect on any of the materials of construction When using compounds in the Freeze Dryer that are hostile to the mater...

Page 19: ...lvents used in the Freeze Dryer may be flammable or hazardous to your health Use extreme caution and keep sources of ignition away from the solvents When using flammable or hazardous solvents the vacu...

Page 20: ...to begin using your Freeze Dryer Read this chapter to learn how to operate the controls understand the display connect samples Do not use the Freeze Dryer in a manner not specified by the manufacturer...

Page 21: ...efrigeration Switch Used to start only the refrigeration module 7 Manual Operation Indicator When lit the green LED indicates the Freeze Dryer is being controlled manually by the operator Each functio...

Page 22: ...lid gasket to remove any dirt and contaminants that could cause a vacuum leak Vacuum grease is not required on the lid gasket to obtain a proper vacuum seal 4 Remove the accessory drying chamber or m...

Page 23: ...Press SELECT until the desired vacuum operating level is displayed See Setting the Operating Vacuum Level for guidelines If 0 is selected the vacuum control is disabled and the vacuum level in the Fre...

Page 24: ...press SELECT The display will show If you want to reset the service hours to 0 continue to hold SELECT five seconds until hours are reset Press SELECT The display will show the RS 232 data transmissi...

Page 25: ...um is between 0 450 and 0 133 mBar the lower green vacuum graph LED will flash When the system vacuum level is 0 133 mBar the green LED will be lit steadily indicating that most samples may be added M...

Page 26: ...7 Fungi 40 and Colder 50 Colder 0 04 and lower Vegetable Tissue 25 to 50 35 to 60 0 22 to 0 01 Human Tissue 30 to 40 40 to 50 0 12 to 0 04 Blood Plasma 10 to 25 20 to 35 1 03 to 0 22 Vaccine 30 to 40...

Page 27: ...quired connect the gas supply line to the vent port on the valve before turning the plastic knob on the valve to vent position The sample container may now be removed In the vent position the bevel on...

Page 28: ...cal damage to the collector chamber and collector coil will result Do not attempt to chip ice from the collector coil as damage may occur to the coil Never attempt to start the vacuum pump when there...

Page 29: ...When an alarm occurs press MENU and the display will show The alarm will self cancel when the proper voltage is restored The high and low alarm points are preset at the factory to correspond to the n...

Page 30: ...the operation of the Freeze Dryer The pump oil should be regularly monitored to verify that it is clean Moisture in Collector If there is liquid in the collector chamber and either a REFRIGERATION but...

Page 31: ...bove MIN add oil to proper level 4 If oil shows cloudiness particles or discoloration drain the pump and replace with fresh oil 5 Utilization of acids requires immediate cleaning and neutralization af...

Page 32: ...and polishes may be used on the exterior surfaces Do not use solvents to remove stains from the exterior surfaces as they may damage the finish Annually 1 Every 12 months or more often if the Freeze...

Page 33: ...a connecting cable below 1 Computers with a 25 pin D sub male serial connector should use Connect Cable Labconco part number 7537801 to connect a computer to the Labconco Freeze Dryer 2 Computers wit...

Page 34: ...ure not available on 2 5L models PG Purge not available on 2 5L models There are several commercially available software packages which can read RS232 data and enter the data into a computer program s...

Page 35: ...ode and phone number will appear Enter the appropriate numbers and continue 3 The Connection Description dialogue box will open Type in a user defined name and select an icon for the new connection Pr...

Page 36: ...33 8811 or 1 800 522 7658 31 5 The Com X Properties dialogue box will open Enter the appropriate data properties and press OK 6 When the Freeze Dryer main power switch is on the data will be transmitt...

Page 37: ...trical connection of the vacuum pump to the Freeze Dryer If the vacuum pump has a power switch make sure that it is turned on If the vacuum is not adequate when the vacuum pump is operating proceed wi...

Page 38: ...plug is installed and the tube appears to be in good condition 4 Check all rubber vacuum tubing for signs of deterioration or cracking 5 Check all connections and make sure they are secure and leak ti...

Page 39: ...chamber and a manifold Thrust Washer Valve Body Knob Stem Assembly Valve Body Trim Ring Manifold Assembly Knob Stem Assembly Bulkhead Nut Gasket No grease required Bulkhead Fitting Valve Body Thrust W...

Page 40: ...er or manifold 2 Leave the gasket on and turn a large freeze dry flask upside down to cover the connection port 3 Start the Freeze Dryer and observe the vacuum indication If the vacuum indication is g...

Page 41: ...ttom 2 Start the Freeze Dryer and observe the vacuum indication If the vacuum is good the problem is in the collection chamber If it is bad the problem is in the plumbing connections from the collecti...

Page 42: ...ystem has a compressor condenser capillary tube metering device and an evaporator During the operation the evaporator of the high stage system refrigerates the condenser of the low stage system This t...

Page 43: ...pare the readings Temperature offset may be varied 5 C in 1 C increments Vacuum offset may be varied 0 020 mBar in 0 001 mBar increments Line Voltage Alarm Points The Freeze Dryer is designed to alarm...

Page 44: ...Master Reset because all offsets and configurations that have been entered will be lost If you must perform a Master Reset hold SELECT approximately 5 seconds This will reset Vacuum units to mBar Vac...

Page 45: ...if 3 is entered into the display the low line alarm will alarm when the voltage drops to approximately 194V for example 206 4x3 194V There will be no alarm indicated if DISABLED is selected Press MEN...

Page 46: ...to change the status Press MENU and the display will show If you want to reset Total Hours press and hold SELECT until an audible beep is heard The display will default to the main freeze dry display...

Page 47: ...00 522 7658 42 In order to store offsets in memory you Must Press and Hold MENU until display shows Wait approximately 10 seconds and the display will default to the main freeze dry display informatio...

Page 48: ...7658 43 APPENDIX A FREEZE DRYER COMPONENTS The following pages list components that are available for your Freeze Dryer The parts shown are the most common replacement parts If other parts are require...

Page 49: ...l 8 7445500 Printed Circuit Board 9 7728000 Drain Plug 10 1643600 O Ring 11 7645904 Drain Hose 12 7373450 Vacuum Hose Not Shown 13 1336400 Power Cord 115V 60Hz Not Shown 1336500 Power Cord 230V 50Hz E...

Page 50: ...Appendix A Freeze Dryer Components Product Service 816 333 8811 or 1 800 522 7658 45 Model 76705xx 1 2 17 18 9 10 14 3 4 5 6 11 8 7...

Page 51: ...e Stage 7591806 Compressor 230V 50 Hz Single Stage 7591807 Compressor 230V 60 Hz Single Stage 6 1302300 Switch 7 7426000 Control Panel Label 8 7445590 Printed Circuit Board 9 7728000 Drain Plug 10 164...

Page 52: ...Appendix A Freeze Dryer Components Product Service 816 333 8811 or 1 800 522 7658 47 Models 76700xx and 74200xx 5 1 2 14 18 9 10 3 4 17 5 7 8 6...

Page 53: ...Product Service 1 800 522 7658 48 APPENDIX B FREEZE DRYER DIMENSIONS Model 76705xx...

Page 54: ...Appendix B Freeze Dryer Dimensions Product Service 816 333 8811 or 1 800 522 7658 49 Model 76700xx...

Page 55: ...Appendix B Freeze Dryer Dimensions Product Service 816 333 8811 or 1 800 522 7658 50 Model 74200xx...

Page 56: ...50 1 6 14 7670530 230V 198 254 50 1 2 5 7 7670531 230V 198 254 50 1 2 5 7 7670540 230V 187 253 60 1 2 5 7 7670541 230V 187 253 60 1 2 5 7 7670560 230V 198 254 50 1 2 5 7 7670561 230V 198 254 50 1 2 5...

Page 57: ...1 10 Environmental Conditions Indoor use only Maximum altitude 6562 feet 2000 meters Ambient temperature range 41 to 104 F 5 to 40 C Maximum relative humidity 80 for temperatures up to 88 F 31 C decre...

Page 58: ...ect drive pump 195 liters minute 115 VAC 50 60 Hz single phase Includes Pump Exhaust Filter Cat 1473400 7739403 Vacuum Pump Two stage direct drive pump 195 liters minute 230 VAC 50 60 Hz single phase...

Page 59: ...dor Pump Exhaust Filter package of 5 Fits pump exhaust filter 1473400 or vacuum pumps 1472100 and 7739402 1988000 Vacuum Pump Oil 1 Liter A molecularly distilled hydrocarbon oil with low vapor pressur...

Page 60: ...e and solvent Cools to approximately 75 C 7522800 12 Port Drying Chamber 9 high x 7 7 8 diameter 304 stainless steel with 12 freeze dry valves 7509200 Product Heater For use in chamber 7522800 Provide...

Page 61: ...5 deep Type 304 stainless steel with 48 each tube stems 7509600 Sample Valve Kit Includes neoprene valve body knob and installation parts 7522900 16 Port Drying Chamber 13 high x 138 diameter 304 sta...

Page 62: ...Appendix D Freeze Dryer Accessories Product Service 816 333 8811 or 1 800 522 7658 57...

Page 63: ...Appendix D Freeze Dryer Accessories Product Service 816 333 8811 or 1 800 522 7658 58...

Page 64: ...Appendix D Freeze Dryer Accessories Product Service 816 333 8811 or 1 800 522 7658 59...

Page 65: ...Appendix D Freeze Dryer Accessories Product Service 816 333 8811 or 1 800 522 7658 60 Ampules...

Reviews: