Directional Airflow
Directional airflow plays a critical role in CApture BT Fuming Chamber performance.
During a Humidity or Fuming Cycle, air is recirculated within the sealed chamber. The
Recirculation Blower pulls air in from the top of the chamber, and returns it at the bottom
of the chamber (shown in Fig. 5-1). During the Exhaust or Purge Cycle, the Exhaust
Blower pulls air from inside the chamber through a Pre-Filter and Main Carbon Filter,
located above the top of the chamber. The Exhaust Blower then pushes the filtered air
back into the room via slot openings located in the back, right side of the chamber
(shown in Fig. 5-1).