Chapter 3: Getting Started
Product Service 1-800-522-7658
11
Original instructions
Good Installation
– Picture 1
Good Installation
– Picture 2
Poor Installation
– Picture 3
Two-piece C-clamp coupler supports
exhaust duct diameter to secure weight load
of duct picked up by structural support.
Structural support is connected to C-clamp coupler.
Note:
Duct diameters, duct materials,
and heights change and Labconco
does not carry the structural supports
or guy wires.
Guy wires from 3 sides are supported
off the roof to prevent wind velocities
from deflecting the exhaust duct.
Blower damage can occur without
structural supports and guy wires.
(Please note the crooked exhaust duct
when installed without structural
supports and without guy wires.)
Summary of Contents for 7061110
Page 36: ...Chapter 5 Using Your Blower Product Service 1 800 522 7658 32 Original instructions...
Page 48: ...Chapter 7 Modifying Your Blower Product Service 1 800 522 7658 44 Original instructions...
Page 58: ...Chapter 9 Troubleshooting Product Service 1 800 522 7658 54 Original instructions...
Page 62: ...Product Service 1 800 522 7658 58 APPENDIX B BLOWER DIMENSIONS...
Page 63: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 59 Original instructions...
Page 64: ...Appendix B Blower Dimensions Product Service 1 800 522 7658 60 Original instructions...
Page 72: ...Product Service 1 800 522 7658 68 APPENDIX F TYPICAL PERCHLORIC ACID HOOD INSTALLATION DIAGRAM...