CHAPTER 1: INTRODUCTION 

 

 

 

 

 

 

 

 

  1 

 

Components Shipped   

 

 

 

 

 

 

 

 

 

  1 

 

General Description   

 

 

 

 

 

 

 

 

 

  2 

 

Performance  

 

 

 

 

 

 

 

 

 

 

 

  2 

 

Component Identification   

 

 

 

 

 

 

 

 

  3 

 

CHAPTER 2: INSTALLATION 

 

 

 

 

 

 

 

 

  8 

 

Location  

 

 

 

 

 

 

 

 

 

 

 

 

  8 

 

Electrical Connection  

 

 

 

 

 

 

 

 

 

  8 

 

Water Connection 

 

 

 

 

 

 

 

 

 

 

  8 

 

Drain Connection  

 

 

 

 

 

 

 

 

 

 

  9 

 

NaOH Pump Connection   

 

 

 

 

 

 

 

 

  9 

 

Optional Water Inlet Kit   

 

 

 

 

 

 

 

 

10 

 

     Rear View of Standard RapidStill II   

 

 

 

 

 

11 

 

     Rear View w/Optional Water Inlet Kit Installed 

 

 

 

12 

 

CHAPTER 3: NORMAL OPERATION   

 

 

 

 

 

13 

Standard Operation Procedure   

 

 

 

 

 

 

 

13 

Safety Precautions 

 

 

 

 

 

 

 

 

 

 

14 

 

CHAPTER 4: ROUTINE MAINTENANCE   

 

 

 

 

16 

 

 

CHAPTER 5: SERVICING THE RAPIDSTILL II 

 

 

 

18 

 

Fuse Replacement 

 

 

 

 

 

 

 

 

 

 

18 

 

Removing the Boiler   

 

 

 

 

 

 

 

 

 

18 

 

Heating Element Replacement  

 

 

 

 

 

 

 

19 

 

Condenser Extension Replacement  

 

 

 

 

 

 

19 

 

Timer Replacement   

 

 

 

 

 

 

 

 

 

20 

 

NaOH System Service 

 

 

 

 

 

 

 

 

 

20 

 

CHAPTER 6: TROUBLESHOOTING

 

 

 

 

 

 

 

22 

 

CHAPTER 7: REPLACEMENT PARTS

   

 

 

 

 

 

24 

 
APPENDIX A: DIMENSIONS 

 

 

 

 

 

 

 

 

26 

 

APPENDIX B: SPECIFICATIONS   

 

 

 

 

 

 

27 

 

APPENDIX C: WIRING DIAGRAM  

 

 

 

 

 

 

28 

 
APPENDIX D: REFERENCES   

 

 

 

 

 

 

 

29 

 

Summary of Contents for 6520000

Page 1: ...RapidStill II Models 6520000...

Page 2: ...ours of 7 30 a m and 5 30 p m Central Standard Time Part 6544100 Rev F ECO L123 Warranty Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reaso...

Page 3: ...l II 11 Rear View w Optional Water Inlet Kit Installed 12 CHAPTER 3 NORMAL OPERATION 13 Standard Operation Procedure 13 Safety Precautions 14 CHAPTER 4 ROUTINE MAINTENANCE 16 CHAPTER 5 SERVICING THE R...

Page 4: ...achieve maximum performance If you have questions or concerns do not hesitate to call us at 1 800 821 5525 for assistance WARNING The disposal of substances used in connection with this equipment may...

Page 5: ...ic wastes etc Sample digestions are accomplished using Labconco s Rapid Digestor systems described in the Accessories section of this manual Figure 1 Performance The RapidStill II produces steam by he...

Page 6: ...ly locates the digestion tube against the connection stopper 5 Safety Door The safety door surrounds the digestion tube and is designed to protect the operator The door should always be in the closed...

Page 7: ...the boiler Low and high water level marks are on the boiler flask to ease the inspection and addition operations 15 Distillate Outlet Tube This tube carries distillate from the condenser to the solut...

Page 8: ...Chapter 1 Introduction Product Service 1 800 522 7658 5 Figure 2...

Page 9: ...n for the condenser cooling water DO NOT restrict the discharge from the drain pipe or its hose The RapidStill II may be damaged by high water pressure 24 Flow Restrictor The flow restrictor prevents...

Page 10: ...Chapter 1 Introduction Product Service 1 800 522 7658 7 Figure 3...

Page 11: ...ater should be municipal drinking water grade and of sufficient quality to prevent or minimize scaling of the boiler The feed water should have a minimum supply volume of 1 gallon 3 8 liters per minut...

Page 12: ...tions Restrictions of the drain line may produce a pressure buildup in the RapidStill II resulting in damage to the unit and water leakage NaOH Pump Connection Using either the silicone tubing supplie...

Page 13: ...a manner similar to the water inlet fitting below it DO NOT overtighten the nut on the bulkhead union it may damage the fitting 5 Remove the flow restrictor assembly from the T shaped fitting Discard...

Page 14: ...Chapter 2 Installation Product Service 1 800 522 7658 11 Rear View of Standard RapidStill II Figure 5...

Page 15: ...Chapter 2 Installation Product Service 1 800 522 7658 12 Rear View of RapidStill II with Optional Water Inlet Kit Installed Figure 6...

Page 16: ...3 Turn the Boiler Heater switch to the ON position After approximately 10 minutes or when the water in the boiler is at a steady boil the unit is ready for operation 4 Place the digestion tube with a...

Page 17: ...with any regulations that apply in the user s area concerning the dumping of waste materials in or upon water land or air and to comply with such regulations The RapidStill II should not be operated i...

Page 18: ...t the boiler run dry Do not attempt to refill the boiler when it is dry or low on water and hot Failure to ensure proper water level in the boiler or attempting to fill it when hot may result in the b...

Page 19: ...the boiler heater d Depress NaOH Addition switch and hold it on until all of the caustic has been purged out of the system and into the distillation tube e Remove the digestion tube and dispose of th...

Page 20: ...with glass USE APPROPRIATE PROTECTIVE SAFETY EQUIPMENT WHEN HANDLING CLEANING SOLUTIONS ALWAYS FOLLOW THE CLEANING SOLUTION INSTRUCTIONS REGARDING THEIR USE SAFETY PRECAUTIONS TO BE FOLLOWED AND DISPO...

Page 21: ...fuse with a proper replacement part only DO NOT attempt to replace the fuse with one of a higher amperage rating 4 To reinstall the fuse reverse the above steps Removing the Boiler 1 Unplug the RapidS...

Page 22: ...raight away from the frame 7 Remove the heating element cup from its platform by removing the three screws on the underside of the cup 8 Remove the heating element from the cup by removing the two hea...

Page 23: ...Disconnect the electrical connections on the timer noting the terminals orientation 5 Using the proper wrench loosen the hex nut that secures the timer to the front panel and pull the timer out of th...

Page 24: ...dStill II Product Service 1 800 522 7658 21 7 Thoroughly rinse the tube and remove the pump fittings and clamps Retain the fittings and clamps for use on the new tube 8 Install the new tube by reversi...

Page 25: ...d not plugged in Plug in cord Boiler doesn t fill Boiler Heater switch is on Turn off boiler heater before filling Cooling water off or flow rate insufficient Check flow rate of condenser cooling wate...

Page 26: ...inked or elevated above the drain pipe Drain hose must flow unrestricted to a drain below the drain pipe Boiling flask breaks Flask has gone dry Do not let flask go dry Water added when boiler is dry...

Page 27: ...ask 1000 M L 3 1 6541600 Boiler Head 4 1 6543500 Balance Tee 5 2 1359400 Plastic Check Valve 6 1 1363500 Steam Control Valve 7 1 4411500 Water Fill Valve 8 1 1624000 Tubing Silicone 1 2 x 5 16 inch or...

Page 28: ...02200 Switch DPDT 15 1 6522900 Connection Stopper 16 2 6520600 Cap Gasket GL 18 17 1 6520700 Dispensing Tube NaOH 18 1 6520300 Distribution Head 19 1 6521200 Cap Gasket GL 32 20 1 6543600 Condenser Ex...

Page 29: ...Product Service 1 800 522 7658 26 A Ap pp pe en nd di ix x A A D Di im me en ns si io on ns s Figure 10...

Page 30: ...i ic ca at ti io on ns s Electrical Requirements 115 VAC 15 Amps 60 Hz single phase Heating Element 1200 Watt 115 VAC Boiler Capacity 1 liter total volume 750 ml operating volume Condenser Flow Rate 1...

Page 31: ...Product Service 1 800 522 7658 28 A Ap pp pe en nd di ix x C C W Wi ir ri in ng g D Di ia ag gr ra am m Figure 11...

Page 32: ...ohol in a Mixture by Direct Distillation 1993 Labconco Corporation Kansas City MO Methodology for the Determination of Ammonia in Water by Direct Distillation 1993 Labconco Corporation Kansas City MO...