Labcold IntelliCold Operating Instructions Manual Download Page 14

Labcold

 

I

ntelli

C

old® refrigerators constantly take readings of the air temperature inside the 

refrigerator and averages and analyses the results. This means that if you have a separate 

monitor it may alarm before the fridge controller, but the refrigerator will still alarm very 

shortly afterwards and, if working correctly will do so long before the temperature poses a 

threat to the contents.

Why doesn’t the alarm react instantly?

There is water at the back of the fridge

Your 

Labcold

 refrigerator automatically defrosts and ice and water may be observed at the 

back. Any ice is melted and runs through the drain hole to the condensate removal tray at 

the back of the fridge. Due to the heat produced by the fridge motor this should evaporate 

but may occasionally cause water to be observed in the tray. This is perfectly normal and is 

nothing to worry about.

21

My independent thermometer shows a diff erent temperature to the front of the fridge

Your 

Labcold

 

I

ntelli

C

old® refrigerator has a probe placed at the rear of the fridge and the 

temperature displayed is calculated by analysing and averaging the temperature data 

collected by the probe. A thermometer placed in a diff erent part of the refrigerator may read a 

slightly diff erent temperature especially if it is placed close to the door or at the top or bottom 

of the fridge as these are the areas where the greatest temperature fl uctuations occur.

Do I really have to have my refrigerator calibrated?

The 

I

ntelli

C

old® controller, which monitors your refrigerator, is designed to be accurate, 

however, the DH Green book now advises that a refrigerator storing vaccines should be 

calibrated annually. By calibrating your fridge annually you have documentary proof that the 

temperature displayed on the front of the refrigerator is the same as that inside the chamber. 

This may be important if the performance of the fridge is questioned by an offi  cial body. 

Labcold are a UKAS calibration laboratory No 8898 and are therefore capable of providing 

calibration to ISO 17025 standards. Why not contact our service team at 

service@labcold.

com

 for more information?

My timer seems to lose and gain minutes

This is because the controller does not contain a proper clock but an oscillator which can 

be aff ected by ambient temperature and power fl uctuations. It will still provide enough 

information to show how long there may have been a temperature incursion so there is no 

need to reset the timer as this is perfectly normal. Please be aware if you 

do

 reset the timer 

you may lose data.

Useful Information

Your 

Labcold

 refrigerator comes complete with a one years parts and labour warranty. 

However, if you register your refrigerator you will be entitled to a 

5 year warranty

 and will be 

sent a 

free

 SD card for automatic 24 hour fridge data logging.

*

To register your warranty, please complete the enclosed warranty form or visit our web site 

at 

www.labcold.com

, click on the ‘Warranty Registration’ tab and follow the instructions. Full 

warranty terms and conditions can also be viewed and downloaded from the 

Labcold 

website.

In the unlikely event that there is a problem with your refrigerator, you should contact your 

retailer in the fi rst instance. If you bought your fridge directly from 

Labcold

 you should contact 

our service team on: 

01256 705 580

 or Email them at 

[email protected]

.

When you register your warranty or want to report a performance issue you will need to have 

the following information to hand. Please fi ll this in as soon as you get your fridge so we can 

help you as quickly as possible in the event of a problem.

                   

                  Model Number

                   Serial Number

                   Retailer

                   Date Purchased

the professional choice for medical 

and scientifi c refrigeration...

Labcold Ltd, Cherrywood, Chineham Park, Basingstoke, RG24 8WF, UK    01256 705 555

* 5 year warranty and free SD card available to UK customers only. Outside the UK, warranty is limited to one         

year parts only .SD cards are available only as a chargeable extra.

22

Summary of Contents for IntelliCold

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions...

Page 2: ...irective and remain the user s responsibility to dispose of This refrigerator contains a flammable gas It MUST NOT be located in a room or area with a volume less than 10m3 This refrigerator contains...

Page 3: ...14 Digital Door Lock 15 Caring for your fridge 18 Loading 18 Cleaning 19 Defrosting 19 Trouble Shooting 20 Useful Information 22 IMPORTANT These operating instructions contain information essential to...

Page 4: ...during manufacture and dry thoroughly Ensure your fridge is level by adjusting the feet at the bottom of your fridge For models RLDF G0501A 1010A and 1510A insert shelves by placing clips in the rack...

Page 5: ...ber on the display will descend accordingly ALTHOUGH THE REFRIGERATOR WILL REACH 5oC AFTER AN HOUR APPROX YOU SHOULD LEAVETHE FRIDGE RUNNING FOR 24 HOURS BEFORE FILLINGWITHTEMPERATURE SENSITIVEVACCINE...

Page 6: ...asons for this such as extremely cold items being placed in the fridge The cause of the drop in temperature should be investigated immediately and the refrigerator should be monitored to check that it...

Page 7: ...uter and or print hard copy The information stored in the controller will cover the last 13 months and is not deleted from the fridges memory when downloaded It is recommended you perform a download o...

Page 8: ...n you have the correct day The display will then change to allow you to set the time Firstly you need to set the hour by using the buttons on the right of the display Press the tick button when you ha...

Page 9: ...s all the temperature data copied to it since the timer was first set on the refrigerator or the last time the data was downloaded from the controller The Labcold SD card can now be inserted into a SD...

Page 10: ...ntinually displays the first two digits of the model number ie 02 with no decimal point or the display does not change to an alarm state or show a temperature value after a few minutes there is a syst...

Page 11: ...around LEDS Red Blue Place the battery terminals on the correct contact points PP3 Battery NOT included Opening the door Once the refrigerator has power you can now open the door using the factory set...

Page 12: ...ion is paid to the loading of the fridge and it is regularly cleaned and defrosted If you are storing vaccines in your IntelliCold pharmacy refrigerator this should be recorded in accordance with reco...

Page 13: ...vaccines If you press the set button on the front it will show the temperature the fridge is set to Should you wish to change the set point please contact the Labcold service team at service labcold...

Page 14: ...idge is questioned by an official body Labcold are a UKAS calibration laboratory No 8898 and are therefore capable of providing calibration to ISO 17025 standards Why not contact our service team at s...

Page 15: ...www labcold com...

Reviews: