background image

Before plugging in your 

Labcold

 refrigerator for the first time make sure the power cable is free 

from damage and check that your voltage corresponds to that on the serial plate. 

It is recommended that you do not use extension cords or multiple socket adapters. Make sure 

the power cable is not bent, pinched or compressed and that it can be easily accessed.

When you first plug your refrigerator in it will alarm and all the icons/buttons will be illuminated.

4

Plug In and Switch On

Setting the Date and Time

This needs to be done as soon as the refrigerator is plugged in. Your fridge does not have a 

GPRS/wireless connection, so this will need to be repeated if the refrigerator experiences a 

power loss and no batteries are fitted. We recommend fitting batteries to the fridge so it will 

alarm and record internal temperatures should it become disconnected from a mains power 

source.

 

The time should NOT be changed when the clocks change to GMT or BST as this may 

cause stored data loss.

To set the date you need to do the following:

H/L

P

H/L

P

The controller will flash 

Y

 followed by 2 digits, for example 

Y98

. Use the up and down arrows 

on the right hand side to adjust to the current year, so if it is 2019 adjust the digits to 

19

. If 

you hold the arrow it will scroll through the numbers, if you touch the arrow(s) the value will 

increase/decrease by one unit at a time. When the controller is showing the correct year, touch 

the tick icon/button and it will store that setting and then ask you to set the month.

The controller will now flash 

П 

followed by 2 digits, for example 

П98

. Adjust the digits using 

the arrow icons/buttons as above, so if it is June adjust the digits to 

06

. When adjusted, touch 

the tick icon/button as before.

H/L

P

H/L

P

Summary of Contents for IntelliCold RLDF/G 19

Page 1: ...www labcold com IntelliCold Pharmacy Refrigerator Operating Instructions RLDF G 19...

Page 2: ...nts indicates that this refrigerator should not be treated as ordinary waste but must be taken to a special collection centre for the recycling of electrical equipment For more information regarding t...

Page 3: ...tbite or freezing of the skin If the power supply cord is damaged it MUST be replaced by the manufacturer its service agent or similarly qualified persons in order to avoid a hazard Do not use mechani...

Page 4: ...are a UKAS calibration laboratory 8898 We can calibrate your refrigerator AND your independent thermometer We don t just use UKAS traceable equipment we use UKAS methodology and training for assured t...

Page 5: ...probe to2 10 Viewing alarm limits 11 Reviewing the date 11 Viewing Min MaxTemperatures 12 Resetting Min MaxTemperatures 12 Alarm Guide 13 Light Icon Button 14 Door Lock 14 Digital Door Lock 15 Micro S...

Page 6: ...your fridge in the location in which it is to be used Do not place your fridge near a direct heat source or direct sunlight and leave at least 6cm all round the unit for adequate ventilation Failure t...

Page 7: ...bcold repeat alarm kit this is where the wires need to be connected Please see the separate instructions for more information The maximum rating for the connection is 1A 30Vdc DO NOT CONNECT TO MAINS...

Page 8: ...ferent functions Like a mobile phone screen it may not work if you re wearing gloves Where the instructions say touch this means tap the icon button like you would an app on your phone Hold means leav...

Page 9: ...e fridge so it will alarm and record internal temperatures should it become disconnected from a mains power source The time should NOT be changed when the clocks change to GMT or BST as this may cause...

Page 10: ...igits using the arrow icons buttons as above so if it is 21st June adjust the digits to 21 When adjusted touch the tick icon button as before H L P H L P After setting the day you next need to set the...

Page 11: ...ate the icons buttons again put your hand across the controller and they will become active after 2 seconds of appearing ALTHOUGH THE REFRIGERATOR WILL REACH 5oC WITHIN ABOUT AN HOUR YOU SHOULD LEAVET...

Page 12: ...e tick icon button when the SEt screen is displayed and the controller will display the set point temperature Viewing the Set Point H L P C H L P To return to the normal display touch the mute icon bu...

Page 13: ...allows you to change the time interval between the alarm being muted and the alarm sounding again Access the menu by holding the two arrow icons buttons simultaneously and use either one of them to sc...

Page 14: ...h the mute icon buttonorjustleavethecontrolleraloneandthescreenwillreturntothenormaltemperaturedisplay Internal Light H L P H L P H L P H L P Battery Alarm The battery alarm is enabled by default If y...

Page 15: ...s etc Like probe to1 the controller will alarm if it detects that the temperature is lower than 2oC or higher than 8oC where it is placed The temperature detected by this probe will display after the...

Page 16: ...holding the two arrow icons buttons and scroll through the options until you come to screen L t H L P Reviewing the Date To review or change the date time enter the menu by holding the two arrow icon...

Page 17: ...ures the controller will revert to the normal display and the icons buttons cease to be active The temperatures recorded by the two probes may differ because they are taking readings from different pa...

Page 18: ...s will display when probe t2 detects that the temperature has exceeded 8oC Because high temperatures can adversely affect the contents of your fridge it should always be investigated The alarm will al...

Page 19: ...controller as described on page 6 and then touch the light icon button The light will turn on even if the door is shut Please note it will not override any settings programmed by entering the Lt func...

Page 20: ...efrigerator There is also a master override key optional extra that enables you to open the refrigerator in the normal way should you wish to or in an emergency The Labcold digital combination lock is...

Page 21: ...next button After the fourth digit is entered the lock will make a noise signifying it has unlocked You can then rotate the lock handle 90o in a clockwise direction to open the fridge If the red LED...

Page 22: ...cked in the event of a mains power failure or if the user code is forgotten The master key is unique to Labcold digital combination locks Should you wish to use the lock as a digital lock again but ha...

Page 23: ...nload this information as a CSV file The controller will store this information for 18 months but we recommend that you download the data once a week especially as the more data there is to download t...

Page 24: ...f you see a screen with Err an error has occurred so try again If you see Er1 it means initialization has failed please contact our service team on 01256 705 580 or service labcold com and if you see...

Page 25: ...warm soapy water and dry them thoroughly before use Should you require any more drawers please contact our sales department on 01256 705 570 Fitting Instructions Shelf supports To install your drawer...

Page 26: ...inserted with the protruding pointed part of the clip at the bottom and facing outwards of the frame When the drawer is placed in the refrigerator the clips on either side will need to be pushed again...

Page 27: ...t the shelf supports fully engage with the 4 holes on the reverse of the drawer frame Holes on the reverse of the drawer frame designed to engage with the shelf supports fitted to the refrigerator rac...

Page 28: ...ame in to the shelf racking as far as they will go The white clips will not engage fully but they will engage enough to stop the drawer from tipping when fully extended Once the drawer is installed yo...

Page 29: ...trong detergents or bleach Wipe the inside of the fridge in the same solution and dry with a soft cloth Take care to avoid the light fitting and ensure that the drain hole at the back of the fridge is...

Page 30: ...d close to the door or at the top or bottom of the fridge as these are the areas where the greatest temperature fluctuations occur Small accumulations of ice on the cold plate are quite normal and sho...

Page 31: ...r fridge directly from Labcold you should contact our service team on 01256 705 580 or Email them at service labcold com When you register your warranty or want to report a performance issue you will...

Page 32: ...www labcold com...